II. temperature Unit Selection

The PT1 allows you to select the type of units used for temperature

measurements. The following table lists preference options with their

corresponding units.

Mode

Unit Preference

C

Degrees Celsius

F

Degrees Fahrenheit

eSC

This is the escape function. Selecting escape exits temperature

unit selection without saving changes and turns the PT1 off.

To set the preference:

1. Press and release the push button. The LCD will briefly display the

firmware version then the current measurement mode.

2. Immediately push and hold the push button. The display will scroll

through “CAL”, “SOL

SEL

”, “FAC

CAL

”, “ºCºF

T

EMP

”, and “ESC”. Release

the button when “ºCºF

T

EMP

” displays.

3. While the display scrolls through “PUSHn

HLD

” and “ºCºF

T

EMP

”, push

and hold the push button. The display will scroll through “C”, “F” and

“ESC”. Release when the desired unit option displays.

4. “SAVEd” displays indicating the unit preference is saved in memory.

Allow the pen to time out to turn power off.

III. Normal operation

Before you take a reading, make sure the pen is clean, calibrated (see

IV. Calibration) and in the appropriate measurement mode. The sample

solution must also be within the specified measurement range. Keep

all foreign material away from the sample to avoid contamination.

Note: If you cannot dip the pen in the sample solution, pour the

sample into a clean container. If you don’t have a sample container

and need to test a vertical stream of solution, use the scoop.
The following table explains what the LED Indicator Light signals mean

and gives the duration of each signal.

Led Indicator Light Signal

Meaning

duration

Rapid Flashing

Dip pen in solution

6 sec

Slow Flashing

Measurement in process

10-20 sec

Solid Light

Note measurement value

6 sec

CAUtIoN:

To measure solution at the extremes of the specified

temperature range, allow the pen to equilibrate by submerging the cell in

the sample solution for 1 minute prior to taking a measurement.

1. Rinse the pen 3 times by submerging the cell in fresh sample solution

and swirling it around.

2. Remove pen from solution, then press and release the push button.

Firmware version will be displayed, then current measurement mode.

3. Grasp the pen by its case with your fingers positioned between the

display and the pen cap to avoid sample contamination.

4. While the LED flashes rapidly, dip the pen in fresh sample solution so

that the cell is completely submerged. If you do not submerge the cell

in solution before the flashing slows, allow the pen to power off and

retake the reading.

5. While the LED flashes slowly, swirl the pen around to remove

bubbles, keeping the cell submerged. Keep the pen at least 1 inch

(2½ cm) away from sides/bottom of container, if applicable.

6. When the LED turns on solid, remove the pen from solution. The display

will alternate between the measurement and temperature readings.

Note the readings for your records.