DESCRIPTION

This manual describes your Myron L Agri-Meter, tells you how to

use it, and how to keep it working accurately for many years.

Myron L AG-5 and AG6/pH Agri-Meters are compact, light-weight,

self-contained conductivity and conductivity/pH meters. They contain
a built-in sample cell cup and, in the case of the AG6/pH, a replaceable
pH sensor. Both Agri-Meters have dials cailibrated in millimhos. The
AG6/pH dial is also calibrated in pH units. It will quickly determine
the conductivity and pH of almost any solution.

Both are 3.4" x 4.5" x 4.0" (85 x 129 x 126mm) and weigh less

than one pound (.45kg). Agri-Meters are completely self-contained.
pH and conductivity readings are automatically temperature
compensated from 50

°

to 160

°

F ( 10

°

to 71

°

C). They are powered

by a 9 volt transistor battery good for at least 2000 tests, or one year
shelf life.

MODEL

CONDUCTIVITY

pH

RANGE

RANGE

AG-5

0-5 millimhos

AG6/pH

0-5 millimhos

2-12 pH

ACCURACY: (Both Models)

CONDUCTIVITY:

±

2% of full scale

pH:

±

0.2 pH units (AG6/pH only)

REPEATABILITY: (Both Models)

±

1%

The pictures in this manual show the major operating parts of

your Myron L Agri-Meter. Handle your Agri-Meter and identify these
parts to become familiar with it.
IMPORTANT: The pH sensor in your AG6/pH Agri-Meter has a limited
life much like a battery. Sensor longevity is affected both by the types
of solution tested and care given your meter. pH sensors are user
replaceable. See MAINTENANCE for additional information.