Nr s, Jk t e – Altek a 830-KP User Manual
Page 22
41
THERMOCOUPLE SPECIFICATIONS
T/C
TYPE
N
R
S
°C
°F
RANGE ACCURACY
RANGE ACCURACY
300 to 1300
±2.1
572 to 2372
±3.0
100 to 299
±2.3
212 to 571
±3.3
-50 to 99
±2.6
-58 to 211
±4.0
-200 to -49
±5.0
-328 to -57
±8.2
1750 to 1768
±4.2
3182 to 3214
±6.9
950 to 1749
±4.0
1742 to 3181
±6.5
650 to 949
±4.4
1202 to 1741
±7.2
300 to 649
±5.1
572 to 1201
±8.4
1700 to 1768
±4.9
3092 to 3214
±8.0
1050 to 1699
±4.4
1922 to 3091
±7.1
700 to 1049
±4.8
1292 to 1921
±7.8
300 to 699
±5.4
572 to 1291
±8.9
Note: Thermocouple accuracies are based on an 80.00 mV Span
T/C Accuracy for °C is ±(0.05% of 80.00 mV + 1°C)
T/C Accuracy for °F is ±(0.05% of 80.00 mV + 1°F)
Source resolution is 1 °C or °F. Read resolution is 0.1 °C or °F
40
THERMOCOUPLE SPECIFICATIONS
T/C
TYPE
J
K
T
E
°C
°F
RANGE ACCURACY
RANGE ACCURACY
100 to 1200
±1.7
212 to 2192
±2.3
-50 to 99
±1.8
-58 to 211
±2.5
-150 to -49
±2.2
-238 to -57
±3.2
-200 to -149
±2.8
-328 to -237
±4.3
1100 to 1372
±2.2
2012 to 2500
±3.1
0 to 1099
±2.0
32 to 2011
±2.9
-100 to -1
±2.3
-148 to 31
±3.3
-200 to -99
±3.6
-328 to -147
±5.7
200 to 400
±1.7
392 to 752
±2.4
0 to 199
±1.9
32 to 391
±2.7
-100 to -1
±2.4
-148 to 31
±3.5
-200 to -99
±3.5
-328 to -147
±5.6
250 to 1000
±1.5
482 to 1832
±2.0
50 to 249
±1.6
122 to 481
±2.1
-100 to 49
±1.9
-148 to 121
±2.6
-200 to -99
±2.6
-328 to -147
±3.9