Technical data / specifications
Pressure Display: ....................................... PSI, Bar, MPa, Kg/cm2
Temperature Display: ................................. °F or °C
Deep Vacuum Display: .............................. Mikron, mBar, KPa, mmHg
Sensing Resolution: ................................... 1 psi (.07 Bar, .007 MPa, .07 Kg/cm2)
Sensing Accuracy: ...................................... ±1 psi or 1% of reading (whichever is greater)
Working Pressure: ..................................... 0 to 750 psi (52 Bar, 5 MPa, 52 Kg/cm2)
Proof Pressure: .......................................... 1000 psi (70 Bar, 7 MPa, 70 Kg/cm2) [Tolerable pressure
without internal damage]
Refrigerant Temperature Range: ............... -40 to 200°F (-4 bis 93°C)
Operating Temperature: ............................ 32 to 122°F (0 bis 45°C)
Temperature Accuracy: ............................. ±1°F (±0.5°C) between 32 to 160°F (0 to 71°C)
Storage Temperature: ............................... 10 to 120°F (12 to 49°C)
Connections: ............................................. 1/4” SAE ( 7/16“ UNF)
Power: ...................................................... 9V DC battery with optional AC/DC adapter
Battery Life: ............................................... 30-36 hours pressure and temperature mode only (continuous
use), 25-30 hours vacuum and backlight mode (continuous
use)
Auto-Off: .................................................. 15 min. with disabling capability
Special features:
• Low battery indicator
• Displays 65 refrigerants
• Displays pressure, saturated/sensor temperature, superheat/subcool temperature and deep vacuum
• Includes back-lit display for easy reading in darker areas
• 9V DC battery with optional AC/DC adapter
• Free .oating piston seal seat aluminum block with sight glass
• Auto-Off with disabling capability
• Protective rubber boot
• Heavy duty hook folds into back of unit
Warning
Wear Safety Glasses / Wear Gloves.
Keep the manifold in a dry place. Do not allow moisture to enter the unit.
Do not vent refrigerant into the atmosphere.
If eyes come in contact with refrigerant, immediately flush with plenty water. Seek medical
attention immediately.
Battery care & installation
When the battery is connected to the unit but not in use it will last six months.
Note: In order to prolong the life of the 9V battery;
a.) keep auto-off function enabled
b.) if pressure, temperature or deep vacuum is being used for an extended period of time turn the unit
off and turn back on again when reading is required. (Allow 30-45 seconds for the correct reading to
appear.)