Psv solenoid valve, Aalborg instruments, 20 corporate drive – Aalborg PSV Series User Manual
Page 2: 3837 w, Td 1293 revision h
TRADEMARKS
Aalborg
7 is a registered trademark of Aalborg Instruments Inc.
Buna
7 is a registered trademark of DuPont Dow Elastometers.
Kalrez
7 is a registered trademark of DuPont Dow Elastomers.
Viton
7 is a registered trademark of Dupont Dow Elastometers L.L.C.
6-32 UNC (2)HOLES
0.156
3.250
0.500
1.000
3.851
1.177
FLOW
Ø 0.250
COMPRESSION
FITTING
0.938
1.250
1.000
0.688
0.156
0.500
4-40 SOCKET SCREW
SPIDER SPRING
STEM
SEAT-VITON INSERT
ORIFICE
O-RING
BLOCK
CORE
SPIRAL SPRING
COMPRESSION SPRING
O-RING
O-RING
ADJUST. SCREW
VALVE BODY
05-19-2006
GUARD ASSEMBLY
NUT
MAINTENANCE
The solenoid valve consists of 316 and
416 stainless steel, and VITON
7 (or
optional BUNA-N
7, EPR or KALREZ7) O-
rings and seals. No regular maintenance
is required except for periodic cleaning.
Various corrosive gases or liquids may
demand more frequent replacement of O-
rings and seals inside the valve. Be sure
to use an elastomer material, appropriate
for your specific application. Contact your
distributor or Aalborg
7 for optional sealing
materials available.
Set the PSV for maximum flow, and
attempt to flush through in both directions
with a clean, filtered, and neutral gas such
as nitrogen. [Another option for fully open-
ing the valve is to remove the plastic cap
on top of the valve, and turn the set screw
counterclockwise until it stops]. If block-
age is not alleviated, return the unit to
your distributor or Aalborg
7 for servicing.
ƽ It is the users responsibility to determine
if the instrument is appropriate for their
OXYGEN application, and for specifying
O2 cleaning service if required. Aalborg
7
is not liable for any damage or personal
injury, whatsoever, resulting from the use
of this instrument for oxygen.
ƽ If the valve is left in the active mode for
an extended period of time, it may
become warm or even hot to the touch.
Use care in avoiding direct contact with
the valve during operation.
ƽ To protect servicing personnel it is
mandatory that any instrument being
serviced is completely purged and
neutralized of toxic, bacteriologically
infected, corrosive or radioactive
contents.
ƽ Use of the PSV valve in a manner other
than that specified in this instruction
sheet, may impair the protection
provided by the equipment.
AALBORG INSTRUMENTS
w
20 CORPORATE DRIVE
w
ORANGEBURG, NY 10962
w
PHONE: 845.770.3000
w
FAX: 845.770.3010
E
-
w
TOLL FREE IN USA
or
CANADA
: 1.800.866.3837
w
WEB SITE
: www.aalborg.com
PSV
SOLENOID VALVE
Aalborg
7 reserves the right to change designs and dimensions at its sole discretion at any time without notice.
For certified dimensions please contact Aalborg
7
.
TD 1293 Revision H