Automatic power-off function

The automatic power-off function turns power off after approximately 2
minutes of stable display during the object(s) applied. When there is no
object applied, it will turn off after 30 seconds.

Calibration

1. Make sure that there is nothing on the weighing pan, press and hold

the LIGHT/CAL key approximately 3 seconds. The display shows

Cal . Release the key.

2. The scale memorizes the zero point and automatically displays

100.0 g .

3. Place the calibration weight of 100g on the center of the pan within 10

seconds.

4. After approximately 8 seconds, the scale will display

f for

success and

e for failure. Then, the scale automatically turns off.

5. Turn the scale on again to start the operation.
G If calibration fails, check if anything is touching the weighing pan,

there is vibration or wind, and the calibration weight is correct.