CALIBRATION

When to Calibrate

Calibration may be required when the scale is initially installed, if it is

moved a substantial distance, or in accordance with local regulations.
This is necessary because the weight of a mass in one location is
not necessarily the same in another location. Also, with time and use,
mechanical deviations can occur.

The Position of the [UNITS] key for the Calibration

The (k)g model has no

[UNITS] key, but the key
for the calibration is
located under the overlay.



Calibration Using a Calibration Weight

Calibrate the scale with the following calibration weight.

SK-1000WP: 1000g ± 0.1g

SK-2000WP: 2000g ± 0.2g

SK-5000WP: 5000g ± 0.5g

SK-5001WP: 5000g ± 0.5g

SK-10KWP: 10kg

±

1g

SK-20KWP: 20kg

±

2g

SK-30KWP:

30kg

±

2g

1. Enter calibration mode.

Press the [ON/OFF] key to turn the power off.

While pressing both the [RE-ZERO] and [UNITS] keys, press the

[ON/OFF] key. Then, Cal will be displayed.


2.

Zero

calibration

Press

the

[RE-ZERO]

key to calibrate zero. Cal0 will be displayed.

When you have proceeded to the display other than Cal0 , press

the [ON/OFF] key to turn the power off and start over again.

Wait for the stable indicator

to be displayed and press the

[RE-ZERO] key. Calf will be displayed in a few seconds. If the

zero calibration only is to be done, press the [UNITS] key. The scale

will show end and automatically return to the weighing mode.


3.

Span

calibration

Place the weight at the center of the platform for accurate calibration.

Wait for the stable indicator

to be displayed and press the

[RE-ZERO] key. The scale will show end and the scale will

automatically return to the weighing mode.