Function parameter settings, C01- 01 – A&D FS-D User Manual
Page 12

10
6. FUNCTION PARAMETER SETTINGS
To design a platform scale, set the capacity, weighing
unit and other functions according to the specifications
of the scale.
Normally, end users do not have to set these function
parameters and these setting can be sealed together
with the calibration switch.
6-1. The procedure for setting parameters
Do not set “C01-1”. Once set, the indicator cannot get into the parameter
setting procedure again.
When turning the power on and the indicator may display ------- ,
press the
PRINT/ENT
key to proceed to the weighing mode.
See “7. CALIBRATION” as for the location of the calibration (CAL) switch.
1. In the weighing mode, press and hold the calibration (CAL)
switch until
Cal
appears. Then, release the switch.
Press and hold the
0
key for about 5 seconds, then you
can also enter the calibration mode
(when “C01-0” is set).
Press the
ON/OFF
key or (CAL) switch to stop the
setting procedure.
2. Press and hold the
ZERO
key and press the
PRINT/ENT
key. Then the first function number is
displayed.
3. Press the
2
key to enter into the “C02 Capacity”
setting.
4. Press the
PRINT/ENT
key. Then the parameter
stored appears.
5. Enter the parameter value for the capacity using the
10-key pad.
6. Press the
ZERO
key to go to the next function
number “C03 Weighing display unit”.
7. Set C03 “0” (metric unit), or “1” (non-metric unit).
8. The settings C02 and C03 are minimum requirements
to design a platform scale. To finalize the settings, be
sure to press the
PRINT/ENTER
key. Then, the
display shows
end
for a few seconds and
returns to weighing mode.
The
ZERO
key increments the function number as in the step 6 above, but note
that all of the new parameters are not stored until
end
is displayed.
To exit with nothing changed, turn the power off without pressing the
PRINT/ENT
key.
Set the other parameters according to the specifications required.
-
C01-
01
Parameter
Function number
-
C01-
0
0
-
C02-
0
0
-
C02-
01
-
C02-
0
2
-00
Cal
00
-
C03-
0
1
-00
end
00