Alternatively, Scorpion front panel settings may be set manually and ap-
plied during the automated measurements under PATS software as fol-
lows:
q
On the Measurement Window click the Pause button
q
On the Scorpion front panel press the “Clr / Local” button to take
manual control of the Scorpion.
q
Use Scorpion front panel controls to set the desired settings and/or
to perform a new calibration. (Note: Unless a new calibration is per-
formed, the manual settings must be consistent with the calibration
to be applied).
q
Use the “Save Cal File to PC.” (Refer to Appendix A for detailed in-
formation on how to use the “Calibration Specification Files.”
4-4
TEST EQUIPMENT
A power meter, power sensor, and Type N calibration component kit is re-
quired for power level and S-parameters calibrations. See Table 1-2, page
1-12.
TEST EQUIPMENT
OPERATIONS, CALIBRATION