11. PRODUCT
SPECIFICATION
Parameter Resolution
Accuracy
Range
Temp Measurement
Flue Temperature with
probe
1.0ºC/F +2.0ºC +0.3% reading
0-600ºC
32-1112ºF
Inlet Temperature
0.1ºC/F
+1ºC +0.3% reading
0-50ºC/32-122ºF
Pressure
0.01 mbar
+2% of full scale
+150mbar to -150mbar
Gas Measurement
*1
Oxygen 0.1%
+0.2% 0-21%
Carbon Monoxide
(standard: H compensated)
1ppm +20ppm <400ppm
+5%<5000ppm
+10%>5000ppm
0-10,000ppm
Carbon Monoxide
(high range)
0.01%
+
5% reading from 0.1%
to 10%
0 - 10%
Nitric Oxide
(standard)
1ppm +5ppm<100ppm
+5%>100ppm
0-5000ppm
Nitric Oxide
(low range)
1 ppm
+2ppm<30ppm
+5ppm>30ppm
0 -100ppm
Nitrogen Dioxide
1 ppm
+5ppm<100ppm
+10ppm<500ppm
+5%>500ppm
0 - 1000ppm
Sulphur Dioxide
1ppm
+5ppm<100ppm
+5%>100ppm
0-5000ppm
Calculations
Carbon Dioxide
*2
0.1% +0.3% reading
0-99.9%
Losses 0.1%
+1.0% reading
0-99.9%
Efficiency 0.1%
+1.0% reading
0-99.9%
Excess Air
0.1%
+0.2% 0-2885.0%
Temp (Nett)
1.0ºC/F
+2ºC +0.3% reading
0-600ºC/32-1112ºF
CO/CO
2
ratio 0.0001
+0.0001 0-0.9999
Poison Index
0.01%
+0.01 0-99.99
Pre-programmed
Fuels
Natural gas, Town gas, Gascor, Light Oil, Heavy Oil, Propane,
Butane, Anthracite, Coke, Coal, Kinsale Gas