beautypg.com

UEi Test Instruments CO71A User Manual

Page 2

background image

Audible alarm
An audible alarm will sound if 35 PPM CO (or great-
er) is detected. The alarm frequency will increase
with higher concentrations. Check the LCD display
for actual level. While the alarm is sounding the unit
can not be turned off.

Back-Light
To activate the back-light press the ON/OFF key
quickly once. Do not hold down the key or the unit
will switch off.To turn off the back-light press the

ON/OFF key quickly again.
Note: Using the back-light decreases battery life.

LOW BATTERY INDICATION

Low battery indicator

If the battery voltage falls below a pre-set level the
symbol “BAT” appears in the display. Do not use the
instrument when this symbol is present. Replace the
battery immediately.

Troubleshooting
If the display fails to stabilize at around zero within a
few minutes this may indicate an expired sensor or a
sensor that has been exposed to excessive levels of CO.
If the symptom persists seek advice from an approved
service center.

Always check for correct operation before each use. Use
a known source of CO (e.g. a smoking match stick or a
recently extinguished flame) taking care to act in a safe
manner and not to expose the CO71A sensor to extreme
temperatures or CO concentration. If the unit does not
change as expected, return the unit to UEi or an approved
service center. DO NOT test using automotive exhaust as
the level may exceed the instrument range. This will poison
the sensor or decrease sensor life span.

If during use, you detect CO leaks, be sure you know the
appropriate actions to take to protect yourself and others.
Before switching off the instrument always expose it to
fresh air until the reading falls below 5 PPM. Note that the
readings fall much more slowly than they rise so you may
need to wait three minutes or longer.

Maintenance
Calibration
The CO71A must be calibrated annually at a certified
sevice location to maintain accuracy. Electrochemical
sensors are subject to long term drift that is corrected
during calibration. Refer to troubleshooting section of this
manual to check responce to CO. THIS DOES NOT verify
accuracy.

Specifications

No user serviceable parts. Re-calibrate every 12
months. Check for correct operation as described in
this manual before every use.

Concentration
of CO in Air
9 ppm

35 ppm

200 ppm

400 ppm

800 ppm

1,600 ppm

3,200 ppm

6,400 ppm

12,800 ppm

Inhalation Times and Toxic

Symptoms*

The maximum allowable

concentration for short term

exposure in a living area

according to ASHRAE.

The maximum allowable

concentration for

continuous exposure in any

eight hour period industry

according to MIOSHA.

Maximum concentration

allowable in any 15 minute

period according to OSHA.

Slight headaches, fatigue,

dizziness, nausea after 2 to

3 hours.

Frontal headaches within 1

to 2 hours, life threatening

after 3 hours. Maximum

allowable limit in flue gas

according to EPA and AGA.

Dizziness, nausea and

convulsions within 45

minutes. Unconsciousness

within 2 hours. Death within

2 to 3 hours.

Headache, dizziness and

nausea within 20 minutes.

Death within 1 hour.

Headache, dizziness and

nausea within 5 to 10

minutes. Death within 30

minutes.

Headache, dizziness and

nausea within 1 to 3

minutes. Death within 10 to

15 minutes.

Death within 1 to 3 minutes.

*The effects can vary depending on size, age, sex and health. UEi
is not liable for the accuracy of this information.

(Table 1)

039

BAT

Range:

0-999 ppm

Resolution:

1 ppm

Accuracy:

±(3% of reading + 1 ppm)

Audible Alarm:

Preset at 35 ppm

Visual Indication:

LCD and LED

Battery:

9V alkaline (NEDA 1604)

Ambient

Temperature:

32 - 104˚F (0 - 40˚C)

Warranty:

3 year