Connecting to the 3000rs, Caution – Transmille 3000RS User Manual
Page 7
3000RS PRECISION RESISTANCE STANDARD
OPERATION MANUAL
Transmille Ltd.
Page 6
Label Reference
Information
Model Number
Model number reference for product
Serial Number
Unique serial number for product
Operating Notes
This section details operational and environmental considerations for the 3000RS
precision resistance standard. Follow these instructions when operating or storing
the resistance standard.
Connecting to the 3000RS
CAUTION
Do not apply voltages above the maximum ratings detailed below. Excessive voltage can
cause permanent damage and affect the nominal resistance values of the reference.
VALUE
MAXIMUM VOLTAGE RATING
1 Ohm
0.5V
10 Ohms
1V
100 Ohms
10V
1kOhm 10V
10kOhms 50V
100kOhms 100V
1MOhm 100V
10MOhm 300V