Sensor

Hall Effect Sensor

12 VDC current
sinking pulse

Materials

Sensor Body

Polypro, Brass, 316 SS

Rotor

Polypro, PVDF optional

Shaft

Nickel-bound tungsten
carbide,zirconia ceramic
optional

Bearings

Ruby

Range

0.2 - 30 FPS (0.06-9.14M/s)

Acuracy

±

1

% FS

Maximum Pressure

Polypro

175 psi @ 75

°

F (12 bar)

Brass

200 (14 bar)

316 SS

250 psi (17 bar)

Maximum Temperature

Polypro

130

°

F (55

°

C) at 0 psi

Brass, SS

200

°

F (93

°

C)

Power

6-24VDC, 8 mA

Nominal K-factor

11 Hz/FPS (3.6 HzM/s)

Cable

#22 AWG 3-con 18' (6m)

Maximum Cable Run

2,000' (650m)

Flow Range (GPM)

Specifications

General Information

The TX80 Series are turbine-type insertion meters de-
signed for use in pipe sizes 1" to 8". High-quality jewel
bearings and nickel-bound tungsten carbide shaft are
used for maximum life and extreme low friction. Bodies
are machined from solid rod for maximum precision. Low-
flow performance is superior. The rotation of the rotor is
detected by a non-drag Hall-effect sensor. Output is a
pulse-type square wave, which can be sent long dis-
tances (up to 2,000 feet) without a transmitter. This sig-
nal can be connected directly to SeaMetrics controls, as
well as PLC's, counters, and computer cards.

SeaMetrics TX80 meters are ideal for chemical propor-
tioning applications. If no display is required, a simple
divider such as the PD10 provides adjustable pump pac-
ing. For rate and total display, as well as pump pacing,
the FT420 flow indicator can be mounted directly on the
TX80 Series, or remotely on a wall or panel.

The TX80 Series require special fittings, since they are
not depth-adjustable, installation in the fitting ensures
correct depth placement in the pipe. Fittings are avail-
able in PVC, brass, and stainless steel. Sensors are avail-
able in brass, 316 stainless steel, and polypropylene. In
plastic pipe 3"-8", use an IP82 sensor, which is 1.00"
longer than the IP81 to accommodate the larger fittings.

LT-10614A

Page 1 of 4

Features

TX80 Series Flow
Sensor Instructions

Pipe stubs provided

for ease of installation.

Solvent weld or

male NPT

Each fitting individually

calibrated and marked

with K-factor

(Pulses per Gallon)

Fitting automatically

sets probe at correct
depth and orientation

Jewel bearings for

superior low-flow

performance.

Threaded housings

allow for field repair

MODEL

SERI

AL

High level (12 VDC)

signal can be sent

up to 2000+ feet

without transmitter

Meter mount option

for electronics:

Rate/Total

4- 20 mA blind transmitter

Pulse Divider

Battery Rate/Total

Min.

Max.

1" 1-1/2"

2"

3"

4"

6"

8"

0.8

1.9

3.1

6.9

12

27

47

80

190

314

691

1200 2700

4700