TROuBLEShOOTinG

Problem

Probable cause

Try...

Laptop screen reads
”No connection,” but
“Active” light is NOT
blinking

Laptop screen reads
”No connection,” but
“Active” light IS blinking

Flow rate incorrect

No flow rate

DL75 is transmitting data.
Wait two minutes before taking action.

Dead battery in DL75

Loose cable

Bad cable

K-factor setting is incorrect

Damaged flow sensor

Broken wire

Wait for “Active” light to blink

Replace battery

Tighten cable

Replace cable

Make sure K-factor in DL75 matches flow
sensor

Mechanical: See if rotor spins freely
EX Magmeter: See if “Status LED” indicator
is blinking

Check for flow rate