Technical data, 2 technical data – KROHNE OPTIBATCH 4011 C EN User Manual
Page 31
TECHNICAL DATA
7
31
OPTIBATCH 4011 C
www.krohne.com
01/2012 - 4000868402 - MA OPTIBATCH R02 en
7.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local representative.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principle
Coriolis mass flow
Application range
Measurement of liquids in batching and filling machines
Measured values
Mass
Calculated values
Volume
Design
Basic
System consists of a measuring sensor with integral converter to process
the output signal
Features
Fully welded maintenance- free sensor with twin U-shaped measuring tube
Variants
Variants
Variants
Variants
Compact version
Integral converter
Measuring accuracy
Mass
Mass
Mass
Mass
Q
nom
x 0.1 ≤ Q
±0.15% of actual measured flow rate
Q < Q
nom
x 0.1
±0.1% of actual measured flow rate + zero stability
Volume
Volume
Volume
Volume
Q
nom
x 0.1 ≤ Q
±0.2% of actual measured flow rate
Q < Q
nom
x 0.1
±0.15% of actual measured flow rate + zero stability
Zero stability
Zero stability
Zero stability
Zero stability
Stainless Steel
0.005% of nominal flow
Repeatability (at reference condtions)
Repeatability (at reference condtions)
Repeatability (at reference condtions)
Repeatability (at reference condtions)
Filling time
Standard deviation
1.5 s < Filling time ≤ 3 s
0.13%
3 s < Filling time ≤ 5 s
0.07%
5 s < Filling time
0.04%
Reference conditions
Reference conditions
Reference conditions
Reference conditions
Warm-up time
15 min
Product
Water
Temperature
+20°C / +68°F
Operating pressure
1 barg / 14.5 psig
Accreditation
UKAS to EN17025
MA_OPTIBATCH_en_120302_4000868402_R02.book Page 31 Thursday, August 2, 2012 3:17 PM