3.5 Namur

versions

See connection diagram below for standard version with NAMUR outputs (non-MODIS)
See paragraph 4.4 how to set-up the NAMUR NE 43 failure indication

Symbols and specifications

Milliampère meter Ri

680 Ohm


Electronic or electro-mechanical totalizer
U

32 V DC / 24 V AC; I

150 mA


Switch, N/O contact 32 V DC / 1.5 mA

External power supply, DC voltage
For passive current output: V

ext

= 15 – 24 V DC; I

22

mA
For passive digital input: V

ext

= 15 – 32 V DC; I

1.5 mA


External power supply, DC or AC voltage
V

ext

32 V DC /

24 V AC: I

150 mA


External current source, e.g. temperature transducer
I

ext

25 mA


Abbreviations and specifications

A1
P

P
I/C

L, 1L

N, 0L

PE/FE

Earth contact for current output and analogue input
Analogue input : 0 – 20 mA / I

25 mA

Earth contact for pulse output
Pulse output :

32 V DC /

24 V AC: I

150 mA

Current output I : 0 – 22 mA / Rload

680 Ohm / 24 V DC or

Digital input C : 0 – 5 V DC = ‘low’ / 15 – 32 V DC = ‘high’
Life mains supply terminal : 100…240 V AC, 24 V AC or 24 V DC
Neutral mains supply terminal : 100…240 V AC, 24 V AC or 24 V DC
Protective Earth / Functional Earth terminal


All in-/output circuits must be connected in passive mode. Observe the polarity of
the instrument: current (I) always flows towards A1, P and I/C (current sink) !


To connect external devices to the in-/output terminals, the wiring requirements for the specific type
of protection of the terminal compartment (standard: increased safety "e", optional: flameproof
enclosure "d") must be respected, see standard EN 60079-14 or the corresponding national
standard.