Technical Data

The responsibility as to the suitability, intended use and corrosion-resistance of the materials used in
their construction rests solely with the purchaser.

Function and applications
Automatic verification of …

KROHNE electromagnetic flowmeters (EMFs)
with IMoCom signal converters

IFC 010 K+F

IFC 020 K+F+E*

IFC 090 K+F (but not “i”)

IFC 110 F

IFC 110 PF (TIDALFLUX, partially filled,
without testing level measurement)

IFC 210 E*

with all connected primary heads of the IFS series and M 900

*

Please note: special adapter (connection) leads are required for
all systems and signal converters marked with “*” currently not
included with supply, special fabrication.

Manual testing of …

KROHNE signal converters, including older series*,
but excluding testing of primary heads, such as

T 900 F+E

SC 80 A+AS / SC 100 A+AS

IFC 080 K+F / IFC 200 / IFC 200 E / AQUAFLUX 070

all above signal converters with IMoCom interface

Not suitable for …

all systems of hazardous duty (Ex) design *

flowmeters K 300 / K 310 / CAPAFLUX / BATCHFLUX /
ALTOFLUX 2W and all systems of hazardous-duty (Ex) design

all systems with signal converters SC 150 F / F 200 (DELAFLUX) /
TIV 60 / TIV 500 (a.c. field) / IFC 090 K or F of “i” design

Scope of test
with automatic verification

IMoCom signal converter

Field current and field frequency

Primary signal processing (input amplifier,
analog-digital conversion at 25%, 50%, 75%, 100% of
set measuring range)

Active curent output at factory calibration points

Pulse output at 50% of set full scale range

The MagCheck PC program identifies and signals changes to the
set values between verifications that affect measuring accuracy

Primary head:

Electrode resistances

Insulation resistance of the field coils

Resistance of field coils

Manual testing

KROHNE signal converters only

Field current and field frequency
at selectable flow values (0-100% of set full-scale range,
adjustable in steps of 0.1%):

Primary signal processing (input amplifier, analog/digital converter)

Current output

Pulse output
additionally for IMoCom signal converters:

Hardware info

Status info of ADC, display, I/O