KROHNE MAGCHECK EN User Manual
Page 5
MagCheck
5
MagCheck
Technical Data
The responsibility as to the suitability, intended use and corrosion-resistance of the materials used in
their construction rests solely with the purchaser.
Function and applications
Automatic verification of …
●
KROHNE electromagnetic flowmeters (EMFs)
with IMoCom signal converters
●
IFC 010 K+F
●
IFC 020 K+F+E*
●
IFC 090 K+F (but not “i”)
●
IFC 110 F
●
IFC 110 PF (TIDALFLUX, partially filled,
without testing level measurement)
●
IFC 210 E*
●
with all connected primary heads of the IFS series and M 900
*
Please note: special adapter (connection) leads are required for
all systems and signal converters marked with “*” currently not
included with supply, special fabrication.
Manual testing of …
●
KROHNE signal converters, including older series*,
but excluding testing of primary heads, such as
●
T 900 F+E
●
SC 80 A+AS / SC 100 A+AS
●
IFC 080 K+F / IFC 200 / IFC 200 E / AQUAFLUX 070
●
all above signal converters with IMoCom interface
Not suitable for …
●
all systems of hazardous duty (Ex) design *
●
flowmeters K 300 / K 310 / CAPAFLUX / BATCHFLUX /
ALTOFLUX 2W and all systems of hazardous-duty (Ex) design
●
all systems with signal converters SC 150 F / F 200 (DELAFLUX) /
TIV 60 / TIV 500 (a.c. field) / IFC 090 K or F of “i” design
Scope of test
with automatic verification
IMoCom signal converter
●
Field current and field frequency
●
Primary signal processing (input amplifier,
analog-digital conversion at 25%, 50%, 75%, 100% of
set measuring range)
●
Active curent output at factory calibration points
●
Pulse output at 50% of set full scale range
●
The MagCheck PC program identifies and signals changes to the
set values between verifications that affect measuring accuracy
Primary head:
●
Electrode resistances
●
Insulation resistance of the field coils
●
Resistance of field coils
Manual testing
KROHNE signal converters only
●
Field current and field frequency
at selectable flow values (0-100% of set full-scale range,
adjustable in steps of 0.1%):
●
Primary signal processing (input amplifier, analog/digital converter)
●
Current output
●
Pulse output
additionally for IMoCom signal converters:
●
Hardware info
●
Status info of ADC, display, I/O