4.3 Status messages and diagnostic information

Measurements out of specification

Failure

Maintenance

Message

Message

Message

Message

Description

Description

Description

Description

Action

Action

Action

Action

S01

pH value > pH max
pH value out of range

Consider the measuring range limits, otherwise
select a suitable sensor for the process conditions of
the application.

S02

pH value < pH min
pH value out of range

S03

Temp. value > Temp max
Temperature value out of range

S04

Temp. value < Temp min
Temperature value out of range

Message

Message

Message

Message

Description

Description

Description

Description

Action

Action

Action

Action

F01

Glass resistance > max. MOhm

Replace the sensor

F02

Glass resistance < min. MOhm

Message

Message

Message

Message

Description

Description

Description

Description

Action

Action

Action

Action

M01

Slope < -65 mV/pH

Recalibrate the sensor. For more information refer
to

Calibration with PACTwareTM on page 17

M02

Slope > -50 mV/pH

M03

Zero point < -99 mV

M04

Zero point > +99 mV

M05

Maintenance interval expired