1

2

3

-50˚C

(-58˚F)

50˚C

(122˚F)

40˚C

(104˚F)

-40˚C

(-40˚F)

80˚C

(176˚F)

0˚C

(32˚F)

100˚C

(212˚F)

150˚C

(302˚F)

200˚C

(392˚F)

250˚C

(482˚F)

Fig. 15: Ambient temperature - Product temperature

1

Product temperature

2

Ambient temperature

3

Temperature range with temperature adapter

Density

- Standard

>0.02 g/cm³ (0.0007 lbs/in³)

- adjustable

>0.008 g/cm³ (0.0003 lbs/in³)

Electromechanical data
Cable entry/plug

2)

- Single chamber housing

l

1x cable entry M20x1.5 (cable-ø 5 … 9 mm),

1x blind stopper M20x1.5

or:

l

1x closing cap ½ NPT, 1x blind plug ½ NPT

or:

l

1x plug (depending on the version), 1x blind

plug M20x1.5

Spring-loaded terminals

for wire cross-section up to 1.5 mm²

(0.0023 in²)

Adjustment elements
Mode switch
- min.

Min. detection or dry run protection

- max.

Max. detection or overfill protection

2)

Depending on the version M12x1, according to DIN 43650, Harting, Am-

phenol-Tuchel, 7/8" FF.