4

ELECTRICAL CONNECTIONS

10

DK46 - DK800

www.krohne.com

08/2012 - 4000430903 AD DK46/800 II2G R04 en

Electrical connections

4.1 Electrical connection

The built-in intrinsically safe NAMUR limit switches in the variable area flowmeter DK.../../../..
may only be connected to isolation switching amplifiers with separated intrinsically safe circuits
to DIN 19234 with the following max. values:

Maximum values of the power supply units

When connecting to intrinsically safe circuits, take into consideration the following maximum
values per circuit as a function of the limit switch (see plate on terminal box) for the energy
stores.

4.2 Terminal assignment

Connect the built-in intrinsically safe components to terminals + / - in accordance with the
following illustration and table. Connect one limit switch to each terminal block. Take note of the
polarity indicated.

For devices with just one limit switch, the connection is on terminal 1.

Supply

P

i

[mW]

U

i

[V]

I

i

[mA]

Type 1

34

16

25

Type 2

64

16

25

Type 3

169

16

52

Type of limit switch

C

i

[nF]

L

i

[µH]

RC10-14.-N0.../RC15-14.-N0...

I7R2010-N***/I7R2015-N***

210

100

RC10-14.-N3...

I7R2010-NL***

150

120

RC15-14.-N3...

I7R2015-NL***

150

70

Connection of limit switch - 2 wire NAMUR

1 Lower limit switch to terminal 1
2 Upper limit switch to terminal 2

The connection box includes an EMC filter unit.