Electrical connections, 2 connections – KROHNE OPTIFLUX 2000-4000 IECEx EN User Manual

Page 11

background image

ELECTRICAL CONNECTIONS

3

11

OPTIFLUX

www.krohne.com

06/2010 - 7312092100 - AD OPTIFLUX IECEx R01 en

3.2 Connections

The OPTIFLUX 2100 C / 4100 C flowmeters contain following terminals:

The housing of the OPTIFLUX 2100 C / 4100 C flowmeter must be connected to the equipotential
bonding system of the hazardous area. See IEC 60 079-14 clause 6.3. For this purpose the
internal or external connection facility (PE clamp) can be used.

Terminals

Circuit

Type of protection Ex

L, N, PE

L+, L-

Power

100-230 Vac, +10% / -15%, 8VA, Um = 253 V

or

24 Vac/dc, for connection to a PELV circuit,

AC: +10 / -15%, 8VA

DC: +30% / -25%, 4W

Um = 253 V

Ex e

A+, A, A-

S

C, C-

D, D-

Current output, active or passive

screen (frame)

Status

Pulse

All circuits Un ≤ 32 V, In ≤ 50 mA,

for connection to a PELV circuit, Um = 253 V

Ex e