How to measure, 3 how to measure – KANOMAX 6006 Anemomaster LITE User Manual
Page 11

10
How to Measure
§ 3
How to Measure
- Procedure Flow Chart -
By sliding down the switch, you can change the Time Constant
to 1 second (FAST) or to 5 seconds (SLOW).
When the power is turned off, it will return to the initial setting of
1 second (FAST).
When measurement values fluctuate significantly, setting to
“SLOW” will make the reading easy.
Turn the power on and slide the switch down twice to enter Air
Temperature Measurement Mode.
Do not make a measurement immediately after the mode
change. Especially, when no airflow is present (wind-speed of
0.1m/s or lower), wait for at least 30 seconds before taking a
measurement.
Power is on
Display all items
Air Velocity Measurement
Mode (FAST)
Push and hold the switch for more than 1 second
Slide the switch
down once
Air Temperature
Measurement Mode
Pause
Air Velocity
Measurement Mode
(SLOW)
Slide the switch
up once
Slide the switch
down twice
*When the power is
turned on, air velocity
measurement mode
(FAST) is automatically
selected.
Pause
ON/OFF
Press and hold for
more than 1 second
FAST/SLOW/
Air Temperature
*
Time Constant
determines the time
span of the moving
average. When you
set the Time Constant
to a larger (longer)
value, the readings
will become stable.
When you select a
smaller (shorter)
value, the readings
will become more
responsive and
sensitive to the
change in air velocity.