2. Release the [MODE] key

3. Press [MODE] key again.

4. The LCD will flash “CAL” followed by the required calibration weight

(ie. 300 for the CD-V2-300, or 50 for CD-V2-50). Add the required

weight. The display will show “PASS” then return to normal weighing

mode.

5. Put away cal weight, press [ON/OFF] key to turn off scale, the cali-

bration is finished

** Repeat this process if scale is still displaying incorrect weight.

Please check the availability of weights at the store from which you

have bought your scale.

V. TROUBLESHOOTING:

The primary reasons for inaccuracy or malfunction are low batteries,

incorrect calibration, overload or operating on an unstable surface.

Please keep this in mind and maintain and operate your scale property.

Display

Elimination of errors

[lo] or [88888]

Change the batteries.

[OUt2]

Recalibrate the scale as shown above.

[O-LD]

You are overloading the scale.

[O-LD] or [LLLLL]

The delicate weighing load cells have been

damaged by misusing the scale (i.e. it was

shocked, dropped or overloaded). You can try

recalibrating the scale. If the load cell has not

been damaged too much it will work again after

recalibration.

[UNST]

Use the scale in a more stable position.