I. OPERATION
Place the scale on a flat and level surface. Note: This
scale features a removable protective cover which can
also be used as a weighing tray. To remove the tray,
flip the hinge release switch on the side of the scale.
1. Press the [ I/O ] key.
2. Wait until "0.0g” or "0.00g" is displayed.
3. Place the object(s) to be weighed on the weighing
platform.
4. Using the [ M ] key, you can switch between
weighing units.
II. TARE WEIGHING
Using Tare, you can subtract the weight of a container.
1. Turn the scale on as described above.
2. Place the "tare item" on the platform such as a tray
or container.
3. Press [ T ] and wait until "0.0g" or 0.00g" is
displayed.
4. You can now add items to the tray or container.
III. CALIBRATION
Calibration is NOT required before first use. Only
calibrate if scale is not weighing properly.
off. Calibration is compete!
BT2-500 - Use a 500g calibration weight
BT2-201 - Use a 200g calibration weight
1. Turn the scale on as described above.
2. Once the scale is at zero, press and hold the
[ M ] key for about 3 seconds. The display will show