7. Self-Check & Fine Calibration

Accuracy Self-Check - Plumb point accuracy self-check

1. As shown in figure 1, put the unit on the ground in the direction shown in

figure 1-1.

2. Power on the unit and mark the center of plumb-down as point “E” and

plumb-up point as “A”.

3. Turn the unit 90° (as shown in figure 1-2) and place the down beam on

the “E” mark. Mark the center of the plumb-up point as “B”.

4. Turn the unit 90°again (as shown in figure 1-3) and repeat step 3. Mark

the center of the plumb-up point as “C”.

5. Turn the unit 90°again (as shown in figure 1-4) and repeat step 3. Mark

the center of the plumb-up point as “D”.

6. Marks A-D should all be in the same spot. If not, the unit will need to be

recalibrated. Please see calibration adjustment information.

IMPORTANT: It is the responsibility of the user to verify the
calibration of the instrument before each use.