Setup menu – TIF 3340 Professional Differential Thermometer and InfraRed Thermometer User Manual
Page 4

Set the Sample Rate of
Temperature Data Recording
The default sample rate is one
second, which is displayed on
the secondary display as 0001
(minute–second).
1. While in Setup Menu, press to select
which reading (minute—second) to adjust.
2. Press or until desired setting displays.
3. Press ENTER to store the new setting in
memory.
SETUP MENU
000 1
K
SETUP
Set the Time
The secondary display shows
year, month, and day. The
tertiary display shows hour and
minute. June 10, 2009, 1:31 PM
is displayed in the graphic.
2009 06 10
1331
K
SETUP
1. While in Setup Menu, press to select
which reading to adjust: year—month—day—
hour—minute.
2. Press or until desired setting displays.
3. Press ENTER to store the new setting in
memory.
8888
K
SETUP
Erase the Memory
The number of events stored is
shown on the secondary display.
This procedure outlines how to
clear the memory.
1. While in Setup Menu, press or until
CLR is displayed.
2. Press ENTER.
3. Press or until YES is displayed.
4. Press ENTER to clear the memory.
Back, Sheet 2
Option
Menu Item
Settings
Offset
T1, T2, IR, T3, T4 T1, T2, IR, T3, or T4 offset
Sample Rate rAt
Recording sample rate
Time Setting tiE
Show / set the time
Clear
CLR
Clear memory of events
Sleep Mode
SLP
ON or OFF
Using Offset to Adjust for Probe Errors
The offset option can adjust thermometer
readings to compensate for the errors of a
specific thermocouple and IR temperature. The
adjustment range allowed is: ± 9.0°F or ± 5.0°C.
1. While in Setup Menu, plug the thermocouple
into the input connector.
2. Place the thermocouple in a known, stable
temperature environment (such as an ice dry
well calibrator). Allow the readings to stabilize.
4. Press or until the primary reading
matches the calibration temperature.
3. Press ENTER to store new setting in memory.
Note: The primary display shows the temperature
plus the offset, and the secondary display shows
the offset. Individual offsets can be stored for T1,
T2, IR, T3, or T4.
CHANGING SETUP OPTIONS
1. Press for three (3) seconds to enter
Setup Menu. SETUP appears at the bottom of
the display.
2. Press or to scroll to the desired setup
option.
3. Press ENTER to change this setting.
4. Press or until desired setting displays.
5. Press ENTER to store new setting in memory.
6. Press SETUP for three (3) seconds to exit
Setup Menu.
NOTE: Setup is disabled in MAX MIN AVG mode.