beautypg.com

Troubleshooting – SKC Limited BioLite User Manual

Page 8

background image

Page 6

007-01-002 Issue C

www.skcltd.com

Troubleshooting

Possible Fault

Corrective Action

Pump does not operate

• Check that mains power is present at the mains socket.
• Ensure that pump on/off switch is in the on position.
• Check for blown fuse in the UK 3 pin mains plug and replace with an

identical fuse if required.

• Check for blown inlet fuse at the rear of the pump and replace with an

identical fuse if required.

Pump cannot acheive required
flow rate

• Check for leakages in the sample train.
• Pump vanes damaged or excessively worn. Return pump to SKC

Limited for service / repair.

Pump stops due to over-
temperature

• Check for flow blockages in the sample train and rectify if required. Wait

for pump to cool before restarting.

• Check that the flow adjust control is not fully closed. Wait for pump to

cool before restarting.

• Check that pump cooling fan is running. If cooling fan is not running

return pump to SKC Limited for service / repair.

Pump cooling fan running,
pump not running and
produces ‘humming’ noise.

• Pump rotor vanes have jammed. Switch off pump immediately. Return

pump to SKC Limited for service / repair.