Getting started – SKC Limited Sidekick MTX Pump Operating Instructions User Manual
Page 13

www.skcltd.com
224-SMTXM Issue F
Page 11
Getting Started
4) Flow Adjust Screw (No. 3)
Use the small bladed screwdriver attachment of the supplied toolkit to turn the flow adjust screw. To increase
the flow rate turn clockwise. The span of the flow adjust screw is 12 turns on 224-50MH & 224-51MTX
models, and 15 turns on 224-52MTX models. DO NOT FORCE the flow adjust screw, and make sure the
screwdriver end is securely located in the adjuster screw slot.
APPLICABLE TO MODELS 224-50MH & 224-51MTX ONLY
5) LED Run Indicator (No. 4)
When the pump is started using the on/off switch the LED run indicator (No. 4) will flash on and off to
indicate that the pump is running.
6) Low Battery Shut Down Function
In the event of a low battery condition during a sample run the pump will automatically shut down, and the LED
run indicator is disabled. To recover from a low battery shut down condition fully charge the battery, press the
on/off switch twice to switch the pump off and then on again and the pump will operate as normal.
APPLICABLE TO MODEL 224-52MTX ONLY
7) LCD Screen (No. 1)
The LCD screen displays the elapsed run time in minutes, a low battery status indicator, and a flow fault
indicator.
Flow fault indicator
Low battery indicator
Timer Display