A&D Precision Health Scale UC-321PBT-C User Manual
Page 5

4
Pairing The PBT-C Series Device With a Continua-Compliant Manager
Model UC-321PBT-C is a PBT-C series device. PBT-C series devices are 
designed to be connected only with Continua Managers. 
To pair the device with a Manager, please follow these steps: 
1. Remove the four AA batteries from the device.
2. Discharge the device completely by trying to take a measurement. Press the
“Measurement” switch, tap the toe switch, or tap the surface of the device as 
appropriate. 
Note: the measurement will not proceed and the LCD will remain blank. 
3. Insert the four AA batteries back into the device.
4. Refer to the Manager’s instruction manual to start a Bluetooth discovery and
inquiry process. 
The device will stay discoverable for 60 seconds following battery insertion. 
If the Manager fails to pair with the device, repeat steps 1 through 3. 
5. Once the manager has discovered the device, enter the PIN “123456” as a
password to finish the pairing process. 
When the Manager supports SSP (Secure Simple Pairing), no password is 
required for the pairing process. 
A device that has been certified by Continua displays this 
logo. 
A Bluetooth device has the logo mark of Bluetooth printed 
on it. 
The RoHS (Restriction of Hazardous Substances) 
Directive 
 