beautypg.com

Warning – Perma Pure AS-Series User Manual

Page 2

background image

AMMONIA SCRUBBER SPECIFICATIONS

REPLACING SCRUBBING MEDIA

REPLACEMENT PARTS

When deposits are visible 75% of the way up the
housing, the scrubbing media needs to be replaced
(see Figure 2).

1. Unscrew thumbscrew on bottom of housing.
2. Swing yoke to one side.
3. Separate housing and bottom cap as an assembly

from top cap.

4. Remove spring.
5. Remove old media and dispose of properly (hous-

ing may be rinsed with soapy water to clean).

6. Fill housing with 150cc of Berl Saddles (tap hous-

ing to allow material to settle).

7. Pour 50cc of Scrubbing Media.
8. Replace stainless steel screen on top of media.
9. Replace spring.
10. Clean o-rings on shell and inside top

manifold (replace if necessary).

11. Place center tube into o-ring seal in top cap.
12. Push and twist to seal housing around o-ring.
13. Replace yoke and tighten thumbscrew finger -

tight (do not overtighten).

Part Number

Description

AS-200-08-PSH Stainless steel top & bottom with polysulfone shell
AS-200-08-SSH Stainless steel top & bottom with stainless steel shell
AS-200-3

O-ring replacement set

AS-200-08-EB

15 filling replacement burl saddles and media

The Ammonia Scrubber performance varies
depending on the concentration of ammonia and
flowrate of the gas sample. At a flowrate of 1 lpm
and 1ppm of ammonia, media should last 40,000
hours.

PERFORMANCE

Polysulfone Stainless Steel

Pressure

30 psig

100 psig

Temperature 120°C

200°C*

Port Size

1/4 " NPT

1/4" NPT

Be sure to connect sample inlet to port marked “IN” on

the scrubber housing.

* When thumbscrew is replaced with SST Hex Bolt.

HEATER JACKET SETUP

1

Power Supply Connection for Heater Jacket

1. Connect 120/240VAC power supply to wires on

scrubber (note: white wire must be capped with
wire nut since voltage is present when heater is
energized in 240VAC operation).

2. Allow heater jacket to run for 15-20 minutes

before starting sample flow (heater is equipped
with thermocouple and thermostat to regulate
heater).

Warning!

DO NOT overtighten the thumbscrew.
Doing so may cause the housing to
crack.

Yoke

Thumbscrew

Jacket

Shoulder Screw

Figure 2

Figure 2

YOKE

THUMB SCREW

BURL SADDLES

& MEDIA