Perma Pure Micro-GASS UG User Manual
Page 14

14
Micro-GASS User Manual
APPENDIX C – CAL 3300 Temperature Controller Configuration
1. For all Micro-GASS systems with the CAL3300 Temperature controller, perform the following setup
procedures.
a. Power Up - Apply power to the system and wait for the controller to perform its self test routine.
The display will then be alternately flashing “nonE” and “inPt”.
b. Input sensor type - Press and hold
Ø key. Press p
key repeatedly to select the “tc J” option for
J type thermocouple. Release both keys.
c. Input unit type - Press the
p
key once. The display will then be alternately flashing “unit” and
“nonE”. Press and hold the
Ø key. Press p
key repeatedly to select the “
O
C” unit. Release both
keys.
d. Input setpoint 1 type - Press the
p
key once. The display will then be alternately flashing
“SP1.d” and “nonE”. Press and hold the
Ø
key. Press
p
key once to select the output device
“SSd”. Release both keys.
e. Input the maximum temperature setpoint - Press and hold the
q
key until “LEVL” appears.
Press and hold the
Ø
key. Press the
q
key to move to level 2. Release the two keys. Press and
hold the
p
key until “hi.SC” appears. Press and hold the
Ø
key. Press the
q
key until “65”
appears. Release the two keys.
f.
Input the minimum temperature setpoint - Press the
p
key once. “Lo.SC” appears. Press and
hold the
Ø
key. Press the p
key until “45” appears. Release the two keys.
2. Configure the low temperature alarm.
a. Configure SP2 relay as alarm - Press and hold
p and q
keys simultaneously for 3 seconds to
enter the program mode. Release the two keys. The display will then be alternately flashing
“tunE” and “oFF”. Press the
q
key once. The display will then be alternately flashing “LEVL” and
“1”. Press and hold the
Ø
key. Press the
p key once to move to level 2. Release the two keys.
Press and hold the p
key until “SP2.A” appears. Press and hold the
Ø
key. Press the
p
key
until the display reads FS.Lo. Release the two keys. The display will then be alternately flashing
“SP2.A” and “FS.Lo”.
b. Save configuration into memory - Press and hold
p
and
q
keys simultaneously for 3 seconds.
The display will then be alternately flashing the process temperature and “PArK” indicating that
no setpoint temperature has been entered.