5.0  FUNCTIONS 

5.1 

AUTO

 

POWER

 

OFF 

The auto power off helps conserve power when using on battery or the AC adapter, 
the machine will automatically turn off after 4 min of inactive use or no change in 
the displayed weight.  

6.0  CALIBRATION 

Before calibrating the Dune series it is best to turn on the unit for about 1 min to let 
the components warm up. 

1) Press [on/off] key to turn the power on, if it is not already. 

2) Press and hold the [Unit] key, until “CAL” is displayed 

3) Remove any weight from the pan.  

4) Press the [Unit] key, the display will flash “CAL” them will show the cal mass 

required. If the cal mass is not displayed then the machine is not stable or is 
not  within a zero range.  Turn  off the unit and try  zeroing  the display  before 
trying another calibration.  

5) Place the appropriate calibration mass on the scale: 

Model 

Calibration Mass 

DCT 201 

200g 

DCT 2000 

2000g 

DCT 5000 

5000g 

6) After  the  stable  reading  the  display  will  show  “PASS”  then  revert  to  the 

weighing  mode.  If  the  calibration  does  not  pass  the  display  will  show  “FAIL” 
and shut off.  

7)  The  calibration  is  now  finished.  If  the  calibration  is  correct,  the  display  will 

show the weight value.  

NOTE: If the calibration fails retry, if it still fails then there may be damage to the 
mechanics or an issue with the power. If you do not follow the procedures in a 
timely manner the machine will automatically switch off and the calibration will 
not be accepted.   