Multichannel Systems USB-MEA32-STIM4-System User Manual
Usb-mea32-stim4-system

Technical Specifications
Page 1 of 1
© 2011 Multi Channel Systems MCS GmbH
USB-MEA32-STIM4-System
The USB-MEA32-STIM4-System is an amplifier with integrated
stimulus generator, filter amplification, and analog / digital board
converting analog signals in digital data streams in real time.
Technical Specifications
Operating temperature
10 °C to 50 °C
Storage temperature
0 °C to 50 °C
Relative humidity
10 % to 85 %, non-condensing
Dimensions (L x D x H)
235 x 135 x 55 mm
Weight
900
g
Amplifier
Gain
1000
Bandwidth
0.5
Hz
-
3
kHz
Data
resolution 16
bit
Number of recording channels
32
Input voltage range
+/- 4096 μV
Cross talk (channel to channel)
typical 0.01 %, max. 0.1 %
Data converter and USB interface
Sampling frequency
up to 50 kHz
Maximal data rate
6.4 MByte per second
Data
resolution 16
bit
USB Version
USB 2.0 High Speed
Stimulus Generator
Output current +/-
256
μA
Resolution
13
bit
Maximum frequency
50 kHz
Time resolution
20 μs
PH01 perfusions cannula
Temperature senor type
PT 100 with 4 wire connection
Accuracy
+/-
0.1
°C
Heating element impedance 20
Interface and connectors
USB
USB 2.0 High Speed cable (type A–mini B)
Power Supply
barrel connector 0.7 x 2.35 mm
Ground
common jack 4 mm
Power supply unit (MPU30)
Input voltage
90 – 264 VAC @ 47 – 63 Hz
Output voltage
11 – 13 V
Max.
Power
30
W
Software
Operating system
Microsoft Windows ® 7, XP or Vista
with NTFS
English and German version supported
MC_Rack program
Version 3.6.6 and higher
Data acquisition and analysis software
MC_DataTool program
Version 2.4.3 and higher
Data export
ASCII file (*.txt) binary file (*.raw) format
MC_Stimulus II
Version 3.0.0 and higher
Stimulation program
LTP-Director
Version 1.0.0 and higher
Data acquisition, analysis and stimulation program
Warning: The USB-MEA32STIM4 device may only be used for the specified purpose. Damage of the device and even fatal injuries can result from improper use.
Do not open the data acquisition and stimulation box and do not change hardware configuration as it could lead to improper behavior of the system.