Multichannel Systems 256MEA-SG User Manual
256mea-sg

Multi Channel Systems
MCS GmbH
Aspenhaustrasse 21
72770 Reutlingen
Germany
Fon +49-7121-9 09 25- 0
Fax +49-7121-9 09 25-11
[email protected]
www.multichannelsystems.com
© 2011 Multi Channel Systems MCS GmbH
Product information is subject to change
without notice.
256MEA-SG
256MEA Signal Generator for use with USB-MEA256-System
Control button
LED
DIP switch
Battery
Gold contact pads
256MEA Signal Generator is a convenient tool for USB-MEA256-System users. Use the 256MEA-SG
instead of setting up an experiment with biological sample for training, controlling, and troubleshooting
purposes. This reduces the number of animal experiments and saves laboratory equipment.
Switch on : Press control button. Switch off : Press control button longer than two seconds.
Important : To change DIP switch position, please switch off the device!
!
Table : DIP switch positions, number of control button presses, and corresponding signals
Switch 1
Switch 2
Control button Signal
presses n times
OFF
OFF
MEA-SG ON
Sinus 0.005 Hz
1
Sinus 0.01 Hz
2
Sinus 0.03 Hz
3
Sinus 1.25 Hz
4
Sinus 12.5 Hz
ON
OFF
MEA-SG ON
EPSP
1
Population
Spike
2
Spikes
OFF
ON
MEA-SG ON
ECG Atrium
1
ECG Ventricle
2
Ventricle FP
ON
ON
MEA-SG ON
ERG with Spikes