Multichannel Systems HiClamp User Manual
Hiclamp

Technical Specifications
©2012 Multi Channel Systems MCS GmbH
HiClamp
Technical Specifications
General information
Operating temperature
10 °C to 50 °C
Storage temperature
0 °C to 70 °C
Dimensions (W x D x H)
640 x 600 x 440 mm
Weight 47
kg
Power
230 VAC @ 50 Hz, 115 VAC @ 60 Hz
Amplifier
Type
Integrated TEVC amplifier
Operation
Fully automatic and computer controlled
Reference electrodes
Two independent active bath clamp
Maximal sampling rate
20 kHz
Data resolution
16 bit
Current electrode output
Output range
-105 μA to +105 μA
Effective current resolution
1 nA
Voltage range
-100 V to +100 V
Voltage electrode input
Input range
-500 mV to +500 mV
Voltage resolution
0.015 mV
Clamp voltage setpoint range
-500 mV to +500 mV
Clamp voltage setpoint resolution
1 mV
Typical rise time in voltage clamp mode < 1 ms
Software
HiClamp software
DataMining and DataMerger software
Connection to the computer
USB 2.0 High Speed
Operation system
Microsoft Windows ® 7, Vista or XP with NTFS,
English version
Warning: The device may only be used together with software from Multi Channel
Systems MCS GmbH, and only for the specified purpose. Damage of the device and
even fatal injuries can result from improper use. Do not change hardware
configuration as it could lead to improper behaviour of the system.