Multichannel Systems MEA1060 Amplifiers User Manual
Mea1060 amplifiers
![background image](/manuals/338921/1/background.png)
Technical Specifications
Date of Print: 26.01.2007
MEA1060-Inv-Standard
MEA1060-Up-Standard
MEA1060-Inv-BC-Standard (incl. FA60S-BC) MEA1060-Up-BC-Standard (incl. FA60S-BC)
MEA96-Well-Standard
Operating temperature
10 °C to 50 °C
10 °C to 50 °C
10 °C to 50 °C
10 °C to 50 °C
10 °C to 50 °C
Storage temperature
0 °C to 50 °C
0 °C to 50 °C
0 °C to 50 °C
0 °C to 50 °C
0 °C to 50 °C
Relative humidity
10 % to 85 %
Dimensions (W x D x H)
165 mm x 165 mm x 20 mm (29 mm with
sockets)
165 mm x 165 mm x 14.5 mm (19 mm with
sockets)
165 mm x 165 mm x 20 mm (29 mm with
sockets)
196 mm x 184 mm x 17.5 mm
Weight
800 g
600 g
750 g (without FA60S) + 450 g (FA60S)
750 g (without FA60S) + 450 g (FA60S)
Supply voltage
± 6 VDC to ± 9 VDC
± 6 VDC to ± 9 VDC
± 6.5 VDC to ± 9 VDC
± 6.5 VDC to ± 9 VDC
± 6.5 VDC to ± 9 VDC
Supply current
max. 220 mA, typically ± 150 mA
max. 220 mA, typically ± 150 mA
± 550 mA
± 550 mA
max. 220 mA, typically ± 150 mA
Number of stimulus input channels
60
60
2
2
0
Maximum stimulus input voltage
n/a
n/a
± 5 V
± 5 V
n/a
Number of input channels
60
60
60
60
96
Input voltage
± 4 mV (with respect to the standard gain of
1200)
± 4 mV (with respect to the standard gain of
1200)
± 4 mV (with respect to the standard gain of
1100)
± 4 mV (with respect to the standard gain of
1100)
± 4 mV (with respect to the standard gain of
1000)
Input impedance
10
11
Ω parallel to 10 pF
10
11
Ω parallel to 10 pF
10
11
Ω parallel to 10 pF
10
11
Ω parallel to 10 pF
10
11
Ω parallel to 10 pF
Input noise
< 800 nV
RMS
< 800 nV
RMS
< 800 nV
RMS
< 800 nV
RMS
< 800 nV
RMS
Noise density
e
n
= 15
e
n
= 15
e
n
= 15
e
n
= 15
e
n
= 15
Number of output channels
60
60
60
60
96
Output voltage
± 5 V
± 5 V
± 5 V
± 5 V
± 5 V
Output current
10 mA
10 mA
10 mA
10 mA
10 mA
Output impedance
300
Ω
300
Ω
300
Ω
300
Ω
300
Ω
Bandwidth*
1 Hz to 3 kHz (standard), 1 Hz to 10 kHz
(custom)
1 Hz to 3 kHz (standard), 1 Hz to 10 kHz
(custom)
1 Hz to 3 kHz (standard), 1 Hz to 8 kHz
(custom)
1 Hz to 3 kHz (standard), 1 Hz to 8 kHz
(custom)
0.6 Hz to 2.2 kHz
Filter slope
60 db/decade
60 db/decade
MEA1060-BC-PA: 20 db/decade (high pass), 60
db/decade (low pass); FA60S-BC: 20 db/decade
(high pass), 80 db/decade (low pass)
MEA1060-BC-PA: 20 db/decade (high pass), 60
db/decade (low pass); FA60S-BC: 20 db/decade
(high pass), 80 db/decade (low pass)
60 db/decade (low pass), 40 db/decade (high
pass)
Gain
1200 (standard), 500 to 5000 (custom)
1200 (standard), 500 to 5000 (custom)
1100 (standard), 500 to 5000 (custom)
1100 (standard), 500 to 5000 (custom)
1000
Typical switch enable time
n/a
n/a
20 ns
20 ns
n/a
Heating element (HE)
(separate lid and bottom heating)
Heating temperature
Ambient temperature (min 10 °C) to 50 °C
Ambient temperature (min 10 °C) to 50 °C
Ambient temperature (min 10 °C) to 50 °C
Ambient temperature (min 10 °C) to 50 °C
Ambient temperature (min 10 °C) to 50 °C
Accuracy
0.1 °C
0.1 °C
0.1 °C
0.1 °C
0.1 °C
Recovery time
0.5 min to 2 min
0.5 min to 2 min
0.5 min to 2 min
0.5 min to 2 min
0.5 min to 2 min
Typical time constant
50 s
50 s
50 s
50 s
50 s
Standard hole size
8 mm (HE-Inv-8)
17.6 mm (HE-UP)
8 mm (HE-Inv-8)
17.6 mm (HE-UP-BC)
n/a
Optional hole sizes
12 mm (HE-Inv-12), 23 mm (HE-Inv-23)
-
12 mm (HE-Inv-12), 23 mm (HE-Inv-23)
-
n/a
Input voltage
max 10 V
max 10 V
max 10 V
max 10 V
max 10 V
Input current
max 2 A
max 2 A
max 2 A
max 2 A
max 2 A
Temperature sensor
PT 100
PT 100
PT 100
PT 100
PT 100
Resistance
15 ± 2
Ω
15 ± 2
Ω
15 ± 2
Ω
15 ± 2
Ω
15 ± 2
Ω
Thermal resistance
6 °C / Watt
6 °C / Watt
6 °C / Watt
6 °C / Watt
6 °C / Watt
Calibration constant
T = c * (R-R
0
)/R
0
for c = 259.7 °C
T = c * (R-R
0
)/R
0
for c = 259.7 °C
T = c * (R-R
0
)/R
0
for c = 259.7 °C
T = c * (R-R
0
)/R
0
for c = 259.7 °C
T = c * (R-R
0
)/R
0
for c = 259.7 °C
Microscope compatibility
Inverse microscopes
Upright** microscopes
Inverse microscopes
Upright** microscopes
n/a
Distance focal plane to microscope table
3.5 mm
8 mm
3.5 mm
9.5 mm
MEA1060 Amplifiers
MEA1060 Amplifiers can be ordered with any gain and bandwidth configurations by the user's choice. The name specifies the configuration. The first number after the microscope type specifies the gain, followed by the lowest frequency, and followed by the highest frequency.
For example, MEA1060-Inv-5000-300-3000 stands for a MEA1060 amplifier for inverted microscopes with a gain of 5000, and 300 Hz to 3 kHz bandwidth. MEA1060-BC operates software-controlled and features a blanking circuit that prevents stimulus artifacts.
MEA1060 amplifiers are available for inverted microscopes and upright microscopes, MEA1060-INV and MEA1060-UP, respectively. **= MEA1060-UP amplifiers were optimized for upright microscopes, but can also be used for inverted microscopes, provided that the focal plane is
compatible with the microscope objective. This is useful whenever an excellent accessibility of the MEA from the top is required. With an MEA1060-UP, the MEA culture chamber is the only limiting parameter for accessibility. The opening in the base plate of the amplifier is circular and
hemispherical to follow the dimensions of an objective. The largest opening at the bottom is 35 mm in diameter.
MEA1060 Amplifiers are for use with two-dimensional 60-electrode microelectrode arrays (MEA). MEA 96-Well Amplifier is for use with 96-well plates with integrated recording electrodes (QT-Well plates).
Hz
nV /
Hz
nV /
Hz
nV /
Hz
nV /
Hz
nV /
Page 1 of 1
© 20
14 Multi Channel Systems MCS GmbH