background image

Measurements

Operation Verification

MS278XB OM

5-85

5. Press the Sweep button from the Sweep menu and allow averaging to complete.

The display should look similar to that below:

6. Record the average displayed noise level in Table 5-9 and verify that it meets the specification.
7. Set the Center Frequency of the MS2781B to the next Center Frequency listed in Table 5-9.
8. Press the Sweep button on the Sweep menu and allow averaging to complete.
9. Record the displayed noise level in Table 5-9 and verify that it meets the specification.

10. Repeat steps 7 through 9 for the rest of the center frequencies in Table 5-9.

Figure 5-54. Displaying the Average Noise Level