7440 series, Installation instructions – KING 7440 Series PVC_PVDF User Manual
Page 4
(714) 891-0008 • www.kinginstrumentco.com
When it comes to flow...we’re instrumental.
4
Installation Instructions
Installation Instructions
7440 SERIES ASSEMBLY WITH INDUCTIVE
RING SENSOR
PARTS LIST:
1. Frame Assembly
2. Glass Meter Tube Assembly
3. Sensor Tension Spring
4. Inductive Ring Sensor
5. Alarm Junction Box With Cover
6. Alarm Junction Box Screw
7. Strain Connector
8. Shield
7440 Series
PVC/PVDF
12700 Pala Drive, Garden Grove, CA 92841
www.kinginstrumentco.com / (714) 891-0008
Meters are warranted against defects in materials and workmanship to the
original user for a period of thirteen (13) months from the date of factory
shipment, provided the meter is installed, operated and maintained in
accordance with King Instrument Company's instructions and
recommendations.
This warranty does not apply if failure is caused or contributed to by any of
the following: improper handling, improper storage, abuse, unsuitable
application of the product, lack of reasonable and necessary maintenance,
use exceeding suggested pressure and temperature maximums, improper
packaging for return, or repairs made or attempted to be made by anyone
other than King Instrument Company, Inc.
KING INSTRUMENT COMPANY, INC. MAKES NO WARRANTY
AS TO THE FITNESS OF ITS PRODUCTS FOR SPECIFIC
APPLICATIONS.
This warranty is valid for the original end-user only and does not apply to
products that have been damaged or modified. This warranty is
non-transferrable and is limited to replacement or repair. The liability of King
Instrument Company arising out of its supply of the products, or their use,
shall not in any case exceed the cost of correcting defects in the products as
set forth above.
THIS WARRANTY IS A LIMITED WARRANTY AND SHALL BE IN
LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY
OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. THERE ARE NO OTHER WARRANTIES WHICH
EXIST BEYOND THE DESCRIPTION OR FACE HEREOF.
IN NO EVENT SHALL KING INSTRUMENT COMPANY BE
LIABLE FOR LOSS OF PROFITS, INDIRECT, CONSEQUENTIAL
OR INCIDENTAL DAMAGES.
Products should be returned, prepaid, to King Instrument Company, Inc. with
proof of purchase. Call factory for Return Merchandise Authorization (RMA)
number and return instructions.
2013A
INDUCTIVE RING SENSOR:
All 7440 Series flowmeters may be fitted with one fiber optic sensor.
FIBER OPTIC SENSOR:
The fiber optic sensor is mounted to the frame with a screw in a slotted
screw for adjustability. The sensor uses a pair of fiber optic cables with an
emitter and receiver to transmit the light generated by the sensor. The
sensor can interface directly to a Programmable Logic Controller (PLC) or
when connected to a separate relay can trigger a local or remote alarm.
FIBER OPTIC SENSOR-ELECTRICAL
SPECIFICATIONS
SUPPLY VOLTAGE:
OUTPUT:
OUTPUT RATING:
10-30V DC
NPN Sinking, N.O. & N.C.
150mA max. Total Load
OUTPUT RESPONSE TIME:
1ms
HOUSING RATING:
IP67; NEMA 6
SUPPLY CURRENT:
25mA
PNP Sourcing, N.O. & N.C.
OFF STATE LEAKAGE CURRENT:
1 MICOAMP @ 30V DC
OUTPUT SATURATION VOLTAGE:
<=1V at 10mA DC
>=1.5V at 150mA DC
REPEATABILITY:
0.25ms
OPERATING TEMPERATURE:
-5°F TO +131°F
FRAME ASSEMBLY
GLASS METER TUBE ASSEMBLY
FIBER OPTIC HOUSING SCREW
FIBER OPTIC HOUSING
ALARM JUNCTION BOX SCREW
STRAIN CONNECTOR
FIBER OPTIC CABLE
FIBER OPTIC ADAPTOR
FIBER OPTIC SENSOR
SHIELD
FRAME ASSEMBLY
GLASS METER TUBE ASSEMBLY
INDUCTIVE RING SENSOR
ALARM JUNCTION BOX SCREW
STRAIN CONNECTOR
SHIELD
ALARM JUNCTION BOX WITH COVER
ALARM JUNCTION BOX WITH COVER
All 7440 Series flowmeters may be fitted with one inductive ring sensor. The
inductive ring sensor is mounted on the glass meter tube using springs to put
tension between the frame and glass meter tube. The sensor can be either
proximity or latching type and can only be used with 316SS or Carboloy floats.
The inductive ring sensor produces an electromagnetic field that senses the
metal float within the sensing zone. Inductive ring sensors are 2-wire, DC, low
current devices and are designed to be used with a remote intrinsic safety
barrier / switch isolator.
INDUCTIVE SLOT SENSOR-ELECTRICAL
SPECIFICATIONS
TYPE:
SUPPLY VOLTAGE:
OUTPUT:
OUTPUT LOAD CURRENT:
Inductive Proximity or Latching
5-25V DC (Switch Isolator)
NAMUR
<= 1mA-Float Present
>= 3mA (15mA Max.)-Float Absent
SWITCHING FREQUENCY:
2kHZ
HOUSING RATING:
IP67
WIRING:
PVC covered, 2 Conductor, 26 AWG,
6.5 Ft. Long. Brown (+) / Blue (-)
PEPPERL+FUCHS
General Purpose
Intrinsically Safe
UL:
FM:
SENSOR APPROVALS:
SWITCH ISOLATOR:
Inductive ring sensors are designed to be connected to a switch isolator for
intrinsically safe applications. The purpose of the switch isolator is to supply
electrical signals between safe and hazardous areas in either direction while
limiting the amount of energy that can be transferred even under fault
conditions. Switch isolators are available with 220VAC, 110VAC or 24VDC
supply voltage requirements, contain single pole double throw (SPDT) relays,
and are DIN rail mountable. See switch isolator specifications for electrical
connections and further details.
Intrinsically Safe
CSA:
Intrinsically Safe
CENELEC:
OPERATING TEMPERATURE:
-13°F to 104°F:
-13°F to 158°F:
1
8
" LATCHING
1
8
" PROXIMITY
1
4
" LATCHING
1
4
" PROXIMITY
SENSOR TENSION SPRING
2X
7/11/2013 1:50:17 PM
12700 Pala Drive, Garden Grove, CA 92841
www.kinginstrumentco.com / (714) 891-0008
Meters are warranted against defects in materials and workmanship to the
original user for a period of thirteen (13) months from the date of factory
shipment, provided the meter is installed, operated and maintained in
accordance with King Instrument Company's instructions and
recommendations.
This warranty does not apply if failure is caused or contributed to by any of
the following: improper handling, improper storage, abuse, unsuitable
application of the product, lack of reasonable and necessary maintenance,
use exceeding suggested pressure and temperature maximums, improper
packaging for return, or repairs made or attempted to be made by anyone
other than King Instrument Company, Inc.
KING INSTRUMENT COMPANY, INC. MAKES NO WARRANTY
AS TO THE FITNESS OF ITS PRODUCTS FOR SPECIFIC
APPLICATIONS.
This warranty is valid for the original end-user only and does not apply to
products that have been damaged or modified. This warranty is
non-transferrable and is limited to replacement or repair. The liability of King
Instrument Company arising out of its supply of the products, or their use,
shall not in any case exceed the cost of correcting defects in the products as
set forth above.
THIS WARRANTY IS A LIMITED WARRANTY AND SHALL BE IN
LIEU OF ANY OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY
OR MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. THERE ARE NO OTHER WARRANTIES WHICH
EXIST BEYOND THE DESCRIPTION OR FACE HEREOF.
IN NO EVENT SHALL KING INSTRUMENT COMPANY BE
LIABLE FOR LOSS OF PROFITS, INDIRECT, CONSEQUENTIAL
OR INCIDENTAL DAMAGES.
Products should be returned, prepaid, to King Instrument Company, Inc. with
proof of purchase. Call factory for Return Merchandise Authorization (RMA)
number and return instructions.
2013A
INDUCTIVE RING SENSOR:
All 7440 Series flowmeters may be fitted with one fiber optic sensor.
FIBER OPTIC SENSOR:
The fiber optic sensor is mounted to the frame with a screw in a slotted
screw for adjustability. The sensor uses a pair of fiber optic cables with an
emitter and receiver to transmit the light generated by the sensor. The
sensor can interface directly to a Programmable Logic Controller (PLC) or
when connected to a separate relay can trigger a local or remote alarm.
FIBER OPTIC SENSOR-ELECTRICAL
SPECIFICATIONS
SUPPLY VOLTAGE:
OUTPUT:
OUTPUT RATING:
10-30V DC
NPN Sinking, N.O. & N.C.
150mA max. Total Load
OUTPUT RESPONSE TIME:
1ms
HOUSING RATING:
IP67; NEMA 6
SUPPLY CURRENT:
25mA
PNP Sourcing, N.O. & N.C.
OFF STATE LEAKAGE CURRENT:
1 MICOAMP @ 30V DC
OUTPUT SATURATION VOLTAGE:
<=1V at 10mA DC
>=1.5V at 150mA DC
REPEATABILITY:
0.25ms
OPERATING TEMPERATURE:
-5°F TO +131°F
FRAME ASSEMBLY
GLASS METER TUBE ASSEMBLY
FIBER OPTIC HOUSING SCREW
FIBER OPTIC HOUSING
ALARM JUNCTION BOX SCREW
STRAIN CONNECTOR
FIBER OPTIC CABLE
FIBER OPTIC ADAPTOR
FIBER OPTIC SENSOR
SHIELD
FRAME ASSEMBLY
GLASS METER TUBE ASSEMBLY
INDUCTIVE RING SENSOR
ALARM JUNCTION BOX SCREW
STRAIN CONNECTOR
SHIELD
ALARM JUNCTION BOX WITH COVER
ALARM JUNCTION BOX WITH COVER
All 7440 Series flowmeters may be fitted with one inductive ring sensor. The
inductive ring sensor is mounted on the glass meter tube using springs to put
tension between the frame and glass meter tube. The sensor can be either
proximity or latching type and can only be used with 316SS or Carboloy floats.
The inductive ring sensor produces an electromagnetic field that senses the
metal float within the sensing zone. Inductive ring sensors are 2-wire, DC, low
current devices and are designed to be used with a remote intrinsic safety
barrier / switch isolator.
INDUCTIVE SLOT SENSOR-ELECTRICAL
SPECIFICATIONS
TYPE:
SUPPLY VOLTAGE:
OUTPUT:
OUTPUT LOAD CURRENT:
Inductive Proximity or Latching
5-25V DC (Switch Isolator)
NAMUR
<= 1mA-Float Present
>= 3mA (15mA Max.)-Float Absent
SWITCHING FREQUENCY:
2kHZ
HOUSING RATING:
IP67
WIRING:
PVC covered, 2 Conductor, 26 AWG,
6.5 Ft. Long. Brown (+) / Blue (-)
PEPPERL+FUCHS
General Purpose
Intrinsically Safe
UL:
FM:
SENSOR APPROVALS:
SWITCH ISOLATOR:
Inductive ring sensors are designed to be connected to a switch isolator for
intrinsically safe applications. The purpose of the switch isolator is to supply
electrical signals between safe and hazardous areas in either direction while
limiting the amount of energy that can be transferred even under fault
conditions. Switch isolators are available with 220VAC, 110VAC or 24VDC
supply voltage requirements, contain single pole double throw (SPDT) relays,
and are DIN rail mountable. See switch isolator specifications for electrical
connections and further details.
Intrinsically Safe
CSA:
Intrinsically Safe
CENELEC:
OPERATING TEMPERATURE:
-13°F to 104°F:
-13°F to 158°F:
1
8
" LATCHING
1
8
" PROXIMITY
1
4
" LATCHING
1
4
" PROXIMITY
SENSOR TENSION SPRING
2X
7/11/2013 1:50:17 PM
7440 SERIES FIBER OPTIC SENSOR
FIBER OPTIC SENSOR:
All 7440 Series flowmeters may be fitted with one fiber optic sensor.
The fiber optic sensor is mounted to the frame with a screw in a slotted
screw for adjustability. The sensor uses a pair of fiber optic cables with
an emitter and receiver to transmit the light generated by the sensor.
The sensor can interface directly to a Programmable Logic Control-
ler (PLC) or when connected to a separate relay can trigger a local or
remote alarm.
FIBER OPTIC SENSOR-ELECTRICAL SPECIFICATIONS
SUPPLY VOLTAGE:
10-30V DC
SUPPLY CURRENT:
25mA
OUTPUT:
NPN Sinking, N.O. & N.C.
PNP Sourcing, N.O. & N.C.
OUTPUT RATING:
150mA max. Total Load
OUTPUT RESPONSE TIME:
1ms
OFF STATE LEAKAGE CURRENT: 1 MICOAMP @ 30V DC
OUTPUT SATURATION VOLTAGE: <=1V at 10mA DC
>=1.5V at 150mA DC
REPEATABILITY:
0.25ms
HOUSING RATING:
IP67; NEMA 6
OPERATING TEMPERATURE:
-5°F TO +131°F
7430 SERIES ASSEMBLY WITH
FIBER OPTIC SENSOR
PARTS LIST:
1. Frame Assembly
2. Glass Meter Tube Assembly
3. Fiber Optic Housing Screw
4. Fiber Optic Housing
5. Alarm Junction Box With Cover
6. Alarm Junction Box Screw
7. Strain Connector
8. Fiber Optic Cable
9. Fiber Optic Adaptor
10. Fiber Optic Sensor
11. Shield