A 13- 1, Switch-pro ax13, Specifications – Flowline AX13 Switch-Pro User Manual
Page 2: Ordering dimensions accessories, Strobe alert, Side wall installation installation, Switch-pro, With compact level controller
![background image](/manuals/329572/2/background.png)
Switch-Pro AX13
with Compact Level Controller
®
© 2010 Flowline Inc. • 10500 Humbolt Street Los Alamitos, CA 90720 • 562.598.3015 • www.flowline.com • DSPRAX13 REV A
Select the best level sensor technology based
1)
upon your application.
PFA is available in ultrasonic configuration
2)
only
Specifications
Accuracy:
± 1mm in water
Repeatability:
± 0.5mm in water
LED indication: Sensor, power & relay
Supply voltage: 120 / 240 VAC @ 50-60 Hz.
Consumption: 0.25A maximum
Strobe type:
-X12X: N/A
-X11X: Xenon tube
Strobe flash:
-X12X: N/A
-X11X: (1) per second
Contact type:
(1) SPST relay
Contact rating: 250 VAC @ 10A
Contact output: Selectable NO / NC
Contact delay: 0-60 seconds
Process temp.: F: -40° to 176°
C: -40° to 80°
Ambient temp.: F: -40° to 140°
C: -40° to 60°
Pressure:
150 psi (10 bar) @ 25º
C, derated @ 1.667
psi (0.113 bar) per ºC
above 25º C
Enclosure rating: NEMA 4X (IP65)
Enclosure mat.: -X11X: PP, UL94VO & PC
-X12X: PP, UL94VO
Conduit entrance: 1/2” NPT
Wetted material: AZ13-11XX: Ryton®
AU13-11XX: PP
AU13-21XX: PFA
Process mount: 3/4” NPT (3/4” G)
Mount. gasket: Viton®
Classification: General purpose
Compliance:
CE
Notes
Ordering
Dimensions
Accessories
™
3/4" NPT
(3/4" G)
2.
8"
(7
1
m
m
)
3.2"
(81 mm)
2.1"
(54 mm)
2.8"
(71 mm)
3.9"
(99 mm)
8.2"
(210 mm)
Ordering Notes
1) Select the best sensor technology based
upon your application.
2) PFA is available in ultrasonic configuration.
Sensor technology (1)
Z Vibration
U Ultrasonic
Wetted material (2)
1
PP/Ryton
2 PFA
Configuration
1 Controller with Strobe Alert
2 Controller only
Process mount
0 NPT (US)
4 G (Metric)
6ZLWFK3URZLWK&RPSDFW5HOD\&RQWUROOHU
$B6HULHV
A 13- 1
Switch-Pro
The controller with Strobe Alert configuration flashes
when the relay is energized during a high or low level
alarm condition.
Strobe Alert
The Switch-Pro AX13 series is designed
for installation through the side wall of the
tank. The sensor can be used as a high
or low level alarm as well as overflow pro-
tection or low level pump protection. Use
a tank adapter to make installation quick
and simple.
Side Wall Installation
Installation