Count up timer

Flowchart for Ventilation Pump Failure During Ventilation Cycle

12 (or 24) HOUR CYCLE

STERILIZING

1:53:15 REMAINING

VENTILATING BAG

0:20:12 REMAINING

UNLOAD STERILIZER
VENT PUMP FAILURE

00:02:55

EXIT

VENTILATING BAG

VENT PUMP FAILURE

0:20:01 REMAINING

ALARM OFF

If a VENT PUMP FAILURE occurs during
ventilation cycle, the screen will display the
error message, and beep continuously until
the ALARM button is pressed. The purge
pump will remain on for the remainder of
the normal cycle time and will ventilate the
cabinet.

VENT PUMP FAILURE

REFER TO ERROR

SECTION OF MANUAL

In this case, sterilization has been com-
pleted prior to the failure of the pump.

ITEMS IN THE LINER BAG MAY BE
CONSIDERED STERILE

Refer to the Error and Troubleshooting sections of
the manual for information about repairing the
sterilizer

To reset the machine, shut off the machine power.
Then, while holding both buttons, turn on the
power. Release buttons after the display shows
characters.

The machine may be unloaded once the
counter reaches 0:00:00, and the items may be
used normally

AN74 I

ANPROLENE STERILIZER

START

Press ALARM OFF
to stop the alarm

RESET