FIRST STEPS

1. Connect power lead between CALMASTER and suitable AC mains

supply

2. Switch on AC supply and confirm that CALMASTER red neon is lit
3. Turn all CALMASTER controls to the fully anti-clockwise position
4. IMPORTANT: Remove AC mains lead when testing resistance to

ensure stable resistance reading

Testing AC Voltage

1. Connect ‘Calibrated Meter’ and ‘Meter Under Test’ to CALMASTER left

sockets

2. Switch both ranges to ‘AC Volts’ range
3. Turn ‘AC Volt Adjust’ to the desired voltage and confirm that the

‘Meter Under Test’ reading is within ±5% of the reading displayed on
the ‘Calibrated Meter’

4. Log readings obtained for calibration records
5. Repeat above on all AC voltage ranges on the meters
6. On completion return all control back to the fully anti-clockwise

position

CALMASTER

Operating Instructions

www.actmeters.com

4

www.actmeters.com

5

CALMASTER

Operating Instructions

Testing Resistance

1. Connect ‘Calibrated Meter’ and ‘Meter Under Test’ to the CALMASTER

middle sockets

2. Switch both meters to ‘Resistance’ range
3. Turn the ‘FINE’ or ‘COARSE’ controls to the desired resistance on the

‘Calibrated Meter’

4. Switch the CALMASTER from the ‘Calibrated Meter’ to the ‘Meter

Under Test’ and confirm that the reading is within ±5% of the reading
on the ‘Calibrated Meter’

5. Log readings obtained for calibration records
6. Repeat above on all resistance ranges on the meters
7. On completion return all controls back to the fully anti-clockwise

position

Testing DC Voltage

8. Connect ‘Calibrated Meter’ and ‘Meter Under Test’ to the CALMASTER

middle sockets

9. Switch both meters to ‘DC Volts’ range
10. Switch CALMASTER to the ‘DCV Both Meters’ position
11. Turn the ‘FINE’ or ‘COARSE’ controls to the desired voltage and

confirm that the ‘Meter Under Test’ voltage reading is within ±5% of
the reading displayed on the ‘Calibrated Meter’

12. Log readings obtained for calibration records
13. Repeat above on all DC voltage ranges on the meters
14. On completion return all controls back to the fully anti-clockwise

position