beautypg.com

Vii. composite sampling, Viii. discrete sampling – Xylem WS750 WASTEWATER/STORMWATER SAMPLER User Manual

Page 8

background image

Global Water

800-876-1172

globalw.com

- 8 -

VII. Composite Sampling

A composite sample is a series of smaller samples put into the same bottle, thus showing

an “average” sample over time. Set the Interval control to the desired time between

samples (do not select the External Trigger position). Set the Size control for the

corresponding pump to the desired sample size. Note that the sample size settings are

approximate and apply to a head height of four feet. For heights other than four feet, refer

to the chart inside the front of the sampler. Each pump can have different size settings but

will share the same Sample Interval setting.

The sampler starts timing when both of the optional Rain and Water Sensors detect

moisture. If you do not have the optional water sensors or do not need them for your

application, defeat them by plugging one of the two shorting plugs supplied with the sampler

into their jack. If both inputs are defeated, the sampler starts up triggered when the power

switch is turned on. The Status LED starts blinking every 2 seconds once the sampler is

triggered.

The Start Delay timer begins running when the sampler is triggered by the Rain and Water

Sensors and the first sample is taken when the timer expires. If the Delay control is set to

zero, sampling begins immediately. The delay timer only applies to the first sample taken

by each pump. At the end of each sample, the pumps reverse for 15 seconds.

Sampling continues as Trigger-Delay-Sample-Reverse-Interval-Sample-Reverse-Interval-

Sample-Reverse … until the sample bottles are full. For each pump, the same amount of

Interval time is placed between samples, even if the size settings are different. When both

bottles are full, the Status LED remains lit constantly.

VIII. Discrete Sampling

A discrete sample is a single sample put into the same bottle, thus being a sample at one

point in time. Set the Size control for the corresponding pump to “Full”. Once started, the

pump will continue to run until the sample bottle if full.

As with composite sampling, the Start Delay timer begins when the sampler is trigger by the

Rain and Water Sensors and the first sample is taken when the timer expires. If the Delay

control is set to zero, sampling begins immediately. The Status LED blinks every 2 seconds

while the delay timer or the pumps are running and remains lit constantly once both of the

sample bottles have been filled. At the end of sampling, the pumps reverse for 15 seconds.