11

SDI-12 Command & Response Protocol

Command

Description

aM!

Make Measurement

aMC!

Make Measurement with CRC

aC!

Make Concurrent Measurement

aCC!

Make Concurrent Measurement with CRC

aD0!

Send Data

aR0!

Make Continuous Measurement

aRC0!

Make Continuous Measurement with CRC

aV!

Verify Sensor

aAn!

Change Sensor Address

al!

Send Identification

a!

Send Acknowledge

Standard SDI-12 Commands

Command

Description

aXHELP!

Print a Command Listing

aXRS!

Read Slope

aXWSddd!

Write Slope

aXRO!

Read Offset

aXWOddd!

Write Offset

aXZ!

Zero the Position

aXTD!

Run a Check on the Display

aXSDddd!

Output a Specific Value to the Display

aXTEST!

Initiate a Repeating Test Printout

Extended SDI-12 Commands Unique to the H-3301/3311

The “Measure” command causes a measurement sequence to be performed. Data values generated

in response to this command are stored in the sensor’s buffer for subsequent collection using “D”

commands. The data will be retained in the sensor until another “M”, “C” or “V” command is

executed.

Command

Response

Description

“aM!”

“atttn

Initiate measurement

Where:

a

is the sensor address (“0-9”, “A-Z”, “a-z”, “*”, “?”).

M

is an upper-case ASCII character

Measure Command