beautypg.com

Chapter 2 sdi-12 command and response protocol, H-339 – Xylem H-339 User Manual

Page 13

background image

Chapter 2

SDI-12 Command and Response Protocol

H-339

SDI-12 Command and Response Protocol 2-1

SDI-12 Command and Response Protocol

This is a brief description of the Serial Digital Interface (SDI-12) Command and Response protocol
used by the W

ATER

LOG Series Model H-339 sensor. Included is a description of the commands

®

and data format supported by the H-339.

Refer to the document "A SERIAL DIGITAL INTERFACE STANDARD FOR HYDROLOGIC
AND ENVIRONMENTAL SENSORS". Version 1.2 April 12, 1996 Coordinated by the SDI-12
Support Group, 135 East Center, Logan, Utah.

During normal communication, the data logger sends an address together with a command to the H-
339 sensor. The H-339 then replies with a "response". In the following descriptions, SDI-12
commands and responses are enclosed in quotes. The SDI-12 address and the command/response
terminators are defined as follows:

"a"

Is the sensor address. The following ASCII Characters are valid addresses: "0-9", "A-
Z", "a-z", "*", "?". Sensors will be initially programmed at the factory with the
address of "0" for use in single sensor systems. Addresses "1 to 9" and "A to Z" or
"a to z" can be used for additional sensors connected to the same SDI-12 bus.
Address "*" and "?" are "wild card" addresses which select any sensor, regardless of
its actual address. Unique to the H-339 is that it will always respond to address ‘g’.

"!"

Is the last character of a command block.

""

Are carriage return (0D) hex and line feed (0A) hex characters. They are the
last two characters of a response block.

Notes:

# All commands/responses are upper-case printable ASCII characters.
# Commands must be terminated with a "!" character.
# Responses are terminated with characters.
# The command string must be transmitted in a contiguous block with no gaps of more than

1.66 milliseconds between characters.