2 quality control certification, Advanced instruments inc – Analytical Industries GPR-1800 AIS ATEX Trace PPM Oxygen Analyzer User Manual
Page 3

Advanced Instruments Inc.
3
2 Quality Control Certification
Date:
Customer: Order No.:
Pass
Model:
GPR-1800AIS ppm Oxygen Transmitter S/N __________________
Sensor:
( ) GPR-12-333 ppm Oxygen Sensor
( ) XLT-12-333 ppm Oxygen Sensor
( ) XLT-12-333-F ppm Oxygen Sensor S/N __________________
Accessories:
Owner’s Manual
Configuration:
A-1151-AIS-1 PCB Assembly Main Processing Software Ver: ___________
Power: ( ) A-1166-AIS-AC PCB Assembly Alarms/AC Power 110V & Interconnection
( ) A-1166-AIS-DC PCB Assembly Alarms/DC Power 9-28V & Interconnection
( ) A-1153-AIS-DC PCB Assembly Alarms/DC Power 9-28V & Interconnection
SPDT relays (B-3208, A-3267, Ex-proof window cover)
Ranges: 0-10 ppm, 0-100, 0-1000 ppm, 0-25%
Barometric pressure and temperature compensation
NEMA 4X rated wall mount enclosures
Test: Pass
Pass
Set default zero
Noise level <
± 1.0% FS
Set default span @ 600uA
Alarm delay
Zero calibration
Alarm bypass
Span Calibration
Alarm configurations; ALARM 1, ALARM 2
Analog signal output 4-20mA full scale
Alarm function; ALARM 1, ALARM 2
Calibrates with adequate span adjustment within 10-50% FS
Alarm relays; ALARM 1, ALARM 2
Baseline drift on zero gas <
± 2% FS over 24 hour period
Alarm system fail, dry contact
Overall inspection for physical defects
Options:
( x ) A-3393 H2S Scrubber System
( x ) MRT-1014 LCD Display Low Temperature
( x ) HTR-1004 Heater Low Power Sensor Housing
Notes: