3B Scientific High Current Sensor, 10 A User Manual
3b scientific, Physics
3B SCIENTIFIC
®
PHYSICS
1
High-Current Sensor U11315
Instruction Sheet
09/09 Hh
1. Safety instructions
•
In order to avoid damage to the sensor resistor,
adhere strictly to the current and power limits
mentioned below.
•
Do not exceed a maximum duration of 15 s for
a maximum current of 20 A, while taking read-
ings. Allow the sensor to cool for 10 mins.
•
Use safety connection leads.
2. Description
Sensor box for measuring electric current in AC and
DC circuits.
Shunt resistance measurement method.
For connecting inputs and outputs, use experiment
leads with 4-mm banana plugs.
CAUTION: This sensor box is not automatically
recognised by the interface.
2.1 Contents
1 Sensor box
3. Technical data
Measurement range:
0 to ±10 A
Max. current:
±20 A for 15 s
Sensor type:
Shunt resistance
5 m
Ω / 2 W
Accuracy: <1
%
4. Operation
•
Insert into the current path via the sockets I+
and I-, maintaining the proper polarity for DC.
When measuring AC, the polarity is not signifi-
cant.
•
Set up a connection from the sockets U+ and
U- to the sockets U
A
in
or U
B
in
of the 3B NETlog
TM