higher measurement range to a
lower measurement range.

Before using the device, check the
housing and the measurement
cables for any damage.

Do not conduct measurements in
a damp environment. Workplace,
hands, shoes and floor must be dry.

Before opening the housing all
measurement leads are discon-
nected from the device.

2. Description, technical data

Passive analog multimeter with a
rotary switch for selecting the mea-
surement ranges and a scale with
mirrored background for parallax-
free readings. The device is extreme-
ly robust in terms of load capacity
and is equipped with excellent over-
load protection due to its two anti-
parallel diodes as well as moving coil
movement, which is not sensitive to
external electromagnetic fields. The
safety connection sockets offer pro-
tection against accidental touch con-
tact. The robust plastic housing and
the spring-connected bearing jewels
of the movement guarantee protec-
tion against damage caused by me-
chanical stress.

Measurement ranges:
Voltage measurement:

DC

Meas. range

Int. resistance

100 mV

2 k

1 V

20 k

10 V

200 k

30 V

600 k

100 V

2 M

300 V

6 M

600 V

12 M

AC

Meas. range

Int. resistance

10 V

66,7 k

30 V

200 k

100 V

667 k

300 V

2 M

600 V

4 M

Current measurement:

DC

Meas. range

Voltage drop

50 µA

100 mV

1 mA

500 mV

10 mA

500 mV

100 mA

500 mV

1 A

190 mV

AC

Meas. range

Voltage drop

3 mA

1.5 V

30 mA

1.6 V

300 mA

1.6 V

3 A

1.8 V

Resistance measurement:

Max.

Rotary

Meas. range

meas.

switch

and scale center

current

x 1

1

...35 k

...5 k

45 mA

x 10 10

...350 k

...50 k

4.5 mA

x 100100

...3.5 k

...500 k

0.45 mA

Accuracy: Class 2,5
Influencing variables and nominal
operating ranges:
Temperature 0 – 40° C:
± 1% / 10 K for DC; ± 2.5% / 10 K
for 100 mV/50 µA DC; ± 1.5% / 10 K
for AC
Frequency (30 Hz...1 kHz): ± 2.5%