General notes:
The accuracy of the measurements depends strongly
upon the following factors:
• The precision tube must be extremely clean. If neces-

sary, clean the tube with tissue paper.

• The aluminium cylinder must also be extremely clean.

The slightest amount of dirt such as grease from fin-
gers can lead to considerable friction. Therefore the
cylinder should be cleaned before each measurement
using a fluff-free cloth and some petroleum ether.

• The slightest deformation of the cylinder (e.g. due to

having been dropped) adversely affects the experi-
ment.

• The glass tube must be vertical.

• All stoppers must be airtight.

• Since the duration of the oscillation is squared in the

equation, the time does need to be measured as ac-
curately as possible.

5. Measurement example

Volume V:

10400 cm³

Mass of cylinder m:

15.2 g

Ø

inner

of tube d:

16 mm

Atmospheric pressure p

L

:

1018 mbar

Time t in seconds for five oscillations:

5.172
5.276
5.259
5.224
5.305
5.175
5.231
5.241
5.191
5.175

———

Total: 52.249

Average: 5.2249

Period of oscillation T

s

: 1.04498 s

Substituting into equation (11) gives:

? = 1.39

From published tables:

? = 1.40