Dynatek’s CDT-5 Coating Durability Tester Specifications
CDT-5
Description
Specification
Number of test samples
5
Tube configuration
Straight, curved or bifurcated
Tube length (straight)
88 to ≤ 250 mm
Tube outer diameter
≈1 mm to ≤ 10 mm
Typical distention
0.3% to 18% of tube ID
1
Testing rate
1.2 to 150 Hz
Testing fluid
PBS or distilled water
Fluid temperature
Ambient to ≤ 45º
Flow rate
100 mL/min standard/sample
Test parameters
Stent deflection, tube deflection, pressure, temperature, frequency
Automatic test control method
Tube distention (with optional laser micrometer)
Laser micrometer
Solid-state laser diode (optional; 780 nm, 3 mw)
Pressure transducer range
0-10 psig
Number of particle counters
5
Particle size range
5-1,000 microns
2
Max. particle concentration
15,000 p/mL @ 10 microns
Particle bin range
2-6 bins
System fluid filter
0.2 micron
Particle capture filters
5 (optional)
Particle capture filter size
1 micron (optional, customizable)
Controller
Universal Controller, featuring the NI™ Single-Board RIO platform
Software
Invitroflow software package, with intuitive user
interface to monitor and control the CDT-5
Capacitance tank
0.47 gal
Module
Dimensions
Power
Weight
CDT-5 (base unit)
14 ¾” W X 20”D X 55”H
5.0 A
95 lbs
1
Depending on sample compliance
2
Based on use of appropriate standards
CDT-5™, CDT-20™, Coating Durability Tester™ and Invitroflow™ are trademarks of Dynatek Labs, Inc.