Dynatek Labs CDT-20 User Manual
Page 4
![background image](/manuals/279296/4/background.png)
To receive a customized proposal,
contact us today at:
Dynatek Labs
105 East 4th Street
Galena, MO 65656
800.325.8252
1.417.357.6155
www.dynateklabs.com
Description
Specification
Number of test samples
20
Tube configuration
Straight, curved or bifurcated
Tube length (straight)
Adjustable to ≤220 mm
Tube inner diameter
≈1 mm to ≤15 mm
Typical distention
0.3% to 18% of tube ID
5
Testing fluid
PBS or distilled water
Fluid temperature
Ambient to ≤45º
6
Flow rate
100 mL/min per sample
Test parameters
Stent deflection, tube deflection, pressure, temperature, frequency
Automatic test control method
Tube distention
Pressure transducer
0-10 psig
Laser micrometer
Solid-state laser diode (780 nm, 3 mW)
Number of particle counters
20
Particle size range
5-1,000 microns
7
Max. particle concentration
15,000 p/mL @ 10 microns
Particle bin range
2-6 bins
System fluid filter
0.2 micron
Particle capture filters
20 (optional)
Particle capture filter size
1 micron (optional, customizable)
Controller
Universal Controller, featuring the NI™ Single-Board RIO platform
Software
Invitroflow software package, with intuitive user
interface to monitor and control the CDT-20
Capacitance tank
1.6 gal
Configuration
Dimensions
Weight
Power
Integrated system
38”W x 40”D x 94”H
745 lbs
120 VAC 12 A
Bench-top system
60”W X 72”D X 53”H
600 lbs
120 VAC 12 A
1
US Patent # 7,621,192 B2
2
E.g., coronary or neuro or carotid stent sizes
3
Standards require stent observation to verify pulsatile deflection
4
Customizable to your requirements up to 1,000 microns
5
Depending on sample compliance
6
With optional larger heater
7
Based on use of appropriate standards
Dynatek’s CDT-20 Coating Durability Tester Specifications
CDT-20™ Coating Durability Tester
and Invitroflow™ are trademarks of
Dynatek Labs.